FDA Adverse Event Malfunction Summary report: N

CONNECTING ROD HOFFMANN II MICRO Ø 3 X 90MM

MDR report key: 25317495 · Received May 28, 2026

Report

Report Number
0008031020-2026-00719
Event Type
Malfunction
Date Received
May 28, 2026
Date of Event
October 21, 2025
Report Date
May 28, 2026
Manufacturer
STRYKER GMBH
Product Code
KTT
UDI-DI
04546540389930
PMA / PMN Number
K033476
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT COULD BE CONFIRMED, SINCE THE PRODUCT WAS RETURNED FOR EVALUATION AND MATCHES THE ALLEGED FAILURE. THE DEVICE INSPECTION REVEALED THE FOLLOWING: VISUAL EXAMINATION OF THE RECEIVED PRODUCT SHOWS THAT BOTH ENDS HAVE FINE CRACKS, AS REPORTED. OTHERWISE, THE PART IS IN A GOOD AND NOT DAMAGED CONDITION. INSPECTION OF THE RECEIVED INFORMATION/EVIDENCE ARE DONE OR WILL BE DONE, IN THE PROCEEDINGS OF THE NC. BASED ON INVESTIGATION, THE ROOT CAUSE WAS ATTRIBUTED TO A DESIGN RELATED ISSUE.

Description of Event or Problem · 0

AS REPORTED: "UPON DOING INSPECTIONS FOR PQH 4135808, CRACKS WERE FOUND ON THE ENDS OF ADDITIONAL RODS NOT WITHIN THE SCOPE OF THE PRODUCT HOLD."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
607861 CONNECTING ROD HOFFMANN II MICRO Ø 3 X 90MM APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT KTT STRYKER GMBH J25636 04546540389930

Patients

Seq Age Sex Outcome Treatment
1