FDA Adverse Event
Malfunction
Summary report: N
CONNECTING ROD HOFFMANN II MICRO Ø 3 X 50MM
MDR report key: 25317488
·
Received May 28, 2026
Report
- Report Number
- 0008031020-2026-00720
- Event Type
- Malfunction
- Date Received
- May 28, 2026
- Date of Event
- October 21, 2025
- Report Date
- May 28, 2026
- Manufacturer
- STRYKER GMBH
- Product Code
- KTT
- UDI-DI
- 04546540389916
- PMA / PMN Number
- K033476
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NZ
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
THE COMPLAINT COULD BE CONFIRMED, SINCE THE PRODUCT WAS RETURNED FOR EVALUATION AND MATCHES THE ALLEGED FAILURE. THE DEVICE INSPECTION REVEALED THE FOLLOWING: VISUAL EXAMINATION OF THE RECEIVED PRODUCT SHOWS THAT BOTH ENDS HAVE FINE CRACKS, AS REPORTED. OTHERWISE, THE PART IS IN A GOOD AND NOT DAMAGED CONDITION. INSPECTION OF THE RECEIVED INFORMATION/EVIDENCE ARE DONE OR WILL BE DONE, IN THE PROCEEDINGS OF THE NC. BASED ON INVESTIGATION, THE ROOT CAUSE WAS ATTRIBUTED TO A DESIGN RELATED ISSUE.
Description of Event or Problem · 0
AS REPORTED: "UPON DOING INSPECTIONS FOR PQH 4135808, CRACKS WERE FOUND ON THE ENDS OF ADDITIONAL RODS NOT WITHIN THE SCOPE OF THE PRODUCT HOLD.".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 232087 | CONNECTING ROD HOFFMANN II MICRO Ø 3 X 50MM | APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT | KTT | STRYKER GMBH | V11170 | 04546540389916 |