Report Number
3006897778-2026-00014
Event Type
Malfunction
Date Received
May 27, 2026
Date of Event
April 3, 2026
Report Date
May 27, 2026
Manufacturer
CROSPON LTD
Product Code
FFX
UDI-DI
05391530810050
PMA / PMN Number
K160725
Removal / Correction Number
282497-05-20-2026-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

UPON EVALUATION, ANY CATHETER UTILIZED ON THE EF-100 SYSTEM FOLLOWING (B)(6) 2026 WILL BE INCORRECTLY FLAGGED AS EXPIRED DUE TO THE SYSTEM¿S 24-HOUR USAGE LIMIT, RENDERING THE CATHETER UNAVAILABLE FOR CLINICAL PROCEDURES. THE REAL TIME CLOCK WAS SET UP WITH A BASE YEAR OF 2010 AND WILL NOT SUPPORT CATHETERS AFTER 2025. D10 CONCOMITANT PRODUCT: EF-325N, ENDO CATHETER EF-325N 8 CM NASAL TIP (LOT# 24D1427JZ). MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE CATHETER FAILED THE PRE-USE CHECK THRICE. THE CUSTOMER DISCONNECTED THE CATHETER AND RESTARTED THE SYSTEM, AND THE CATHETER CONTINUED TO FAIL THE PRE-USE CHECK. A NEW CATHETER WAS USED AND IT PASSED THE PRE-USE CHECK. THERE WAS NO PATIENT OR USER HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
352923 ENDOFLIP SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL) FFX CROSPON LTD EF-100 05391530810050

Patients

Seq Age Sex Outcome Treatment
1