FDA Adverse Event Malfunction Summary report: N

ENDOFLIP

MDR report key: 25305803 · Received May 27, 2026

Report

Report Number
3006897778-2026-00002
Event Type
Malfunction
Date Received
May 27, 2026
Date of Event
January 7, 2026
Report Date
May 27, 2026
Manufacturer
CROSPON LTD
Product Code
FFX
UDI-DI
05391530810050
PMA / PMN Number
K160725
Removal / Correction Number
282497-05-20-2026-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

UPON EVALUATION, ANY CATHETER UTILIZED ON THE EF-100 SYSTEM FOLLOWING (B)(6) 2026 WILL BE INCORRECTLY FLAGGED AS EXPIRED DUE TO THE SYSTEM¿S 24-HOUR USAGE LIMIT, RENDERING THE CATHETER UNAVAILABLE FOR CLINICAL PROCEDURES. THE REAL TIME CLOCK WAS SET UP WITH A BASE YEAR OF 2010 AND WILL NOT SUPPORT CATHETERS AFTER 2025. D10 CONCOMITANT PRODUCT: UNK_EF_CATH, UNKNOWN ENDOFLIP CATHETER (LOT#UNKNOWN). MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PROCEDURE COULD NOT BE STARTED AS THERE WAS AN ERROR THAT THE CATHETER HAD BEEN 24 HOURS AND CANNOT BE USED. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
629220 ENDOFLIP SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL) FFX CROSPON LTD EF-100 05391530810050

Patients

Seq Age Sex Outcome Treatment
1