BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET
Report
- Report Number
- 2243072-2026-00401
- Event Type
- Malfunction
- Date Received
- May 27, 2026
- Date of Event
- April 21, 2026
- Report Date
- May 2, 2026
- Manufacturer
- BECTON DICKINSON & CO (FRANKLIN LAKES)
- Product Code
- JKA
- UDI-DI
- 30382903672876
- PMA / PMN Number
- K252506
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE MANUFACTURING LOCATION FOR THIS PRODUCT IS NIPRO. THIS SITE IS AN OEM MANUFACTURING SITE. THEREFORE, BD CORPORATE HEADQUARTERS IN FRANKLIN LAKES, NJ HAS BEEN LISTED IN SECTIONS D.3. AND G.1. AND THE FRANKLIN LAKES FDA REGISTRATION NUMBER HAS BEEN USED FOR THE MANUFACTURE REPORT NUMBER. THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.4. PMA / 510(K)#: K980414. INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 50 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND NO ISSUES WERE OBSERVED RELATING TO SLEEVE FUNCTION AS ALL SAMPLES MET SPECIFICATIONS. ALSO, 8 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING AND NO ISSUES WERE OBSERVED RELATING TO SLEEVE FUNCTION AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE SLEEVE FUNCTION. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT DURING BLOOD DRAW OF ONE PATIENT USING BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET, BLOOD LEAKED FROM THE RUBBER SLEEVE OF ONE DEVICE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 423795 | BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET | TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION | JKA | BECTON DICKINSON & CO (FRANKLIN LAKES) | 24G03A2 | 30382903672876 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |