FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET

MDR report key: 25302616 · Received May 27, 2026

Report

Report Number
2243072-2026-00401
Event Type
Malfunction
Date Received
May 27, 2026
Date of Event
April 21, 2026
Report Date
May 2, 2026
Manufacturer
BECTON DICKINSON & CO (FRANKLIN LAKES)
Product Code
JKA
UDI-DI
30382903672876
PMA / PMN Number
K252506
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE MANUFACTURING LOCATION FOR THIS PRODUCT IS NIPRO. THIS SITE IS AN OEM MANUFACTURING SITE. THEREFORE, BD CORPORATE HEADQUARTERS IN FRANKLIN LAKES, NJ HAS BEEN LISTED IN SECTIONS D.3. AND G.1. AND THE FRANKLIN LAKES FDA REGISTRATION NUMBER HAS BEEN USED FOR THE MANUFACTURE REPORT NUMBER. THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.4. PMA / 510(K)#: K980414. INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 50 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND NO ISSUES WERE OBSERVED RELATING TO SLEEVE FUNCTION AS ALL SAMPLES MET SPECIFICATIONS. ALSO, 8 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING AND NO ISSUES WERE OBSERVED RELATING TO SLEEVE FUNCTION AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE SLEEVE FUNCTION. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING BLOOD DRAW OF ONE PATIENT USING BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET, BLOOD LEAKED FROM THE RUBBER SLEEVE OF ONE DEVICE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
423795 BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION JKA BECTON DICKINSON & CO (FRANKLIN LAKES) 24G03A2 30382903672876

Patients

Seq Age Sex Outcome Treatment
1