I-STAT TROPONIN (HS-TNI ) CARTRIDGE
Report
- Report Number
- 2245578-2026-00118
- Event Type
- Malfunction
- Date Received
- May 27, 2026
- Date of Event
- May 13, 2026
- Report Date
- May 27, 2026
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- MMI
- UDI-DI
- 00054749003051
- PMA / PMN Number
- K240984
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
ON 20-MAY-2026, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT FALSE POSITIVE RESULT ON A PATIENT. THERE WAS NO PATIENT INFORMATION. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD DATE COLLECTED TESTED RESULTS SAMPLE POSITIVE/NEGATIVE I-STAT (B)(6) 2026 10:32 11:48 797.9 NG/L A POSITIVE VITROS (B)(6) 2026 10:32 14:44 <2.25 NG/L A NEGATIVE I-STAT (B)(6) 2026 10:32 14:44 106.7 NG/L A POSITIVE I-STAT (B)(6) 2026 13:40 14:48 20.5 NG/L B NEGATIVE I-STAT (B)(6) 2026 13:40 15:00 44.4 NG/L B POSITIVE VITROS (B)(6) 2026 13:40 15:00 <2.25 NG/L B NEGATIVE I-STAT (B)(6) 2026 15:55 16:00 39.5 NG/L C POSITIVE VITROS (B)(6) 2026 15:55 16:00 <2.25 NG/L C NEGATIVE THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE INVESTIGATION IS UNDERWAY. 99TH PERCENTILE 90% CI SEX N (NG/L, PG/ML) (NG/L, PG/ML) FEMALE 490 13 (10, 17) MALE 404 28 (19, 58) OVERALL 895 21 (14, 30).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 528312 | I-STAT TROPONIN (HS-TNI ) CARTRIDGE | HS-TNI CARTRIDGE | MMI | ABBOTT POINT OF CARE INC. | NA | B25310 | 00054749003051 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |