FDA Adverse Event Malfunction Summary report: N

I-STAT CHEM8+ CARTRIDGE

MDR report key: 25301643 · Received May 27, 2026

Report

Report Number
2245578-2026-00119
Event Type
Malfunction
Date Received
May 27, 2026
Date of Event
May 18, 2026
Report Date
May 27, 2026
Manufacturer
ABBOTT POINT OF CARE
Product Code
JGS
UDI-DI
10054749001917
PMA / PMN Number
K053110
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 0

ON (B)(6) 2026, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGE THAT YIELDED A DISCREPANT CREATININE RESULT ON A 30-YEAR-OLD FEMALE PATIENT WITH LEFT ARM PAIN. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD / DATE / TESTED / RESULTS / SAMPLE: I-STAT, (B)(6) 2026, 17:48, 3.7 MG/DL, A; VITROS 7600, (B)(6) 2026, 19:05, 0.6 MG/DL, B. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
253833 I-STAT CHEM8+ CARTRIDGE CHEM8+ CARTRIDGE JGS ABBOTT POINT OF CARE NA H26085 10054749001917

Patients

Seq Age Sex Outcome Treatment
1