FDA Adverse Event Injury Summary report: N

CONSULTA CRT-D

MDR report key: 2529181 · Received April 11, 2012

Report

Report Number
6000144-2012-02045
Event Type
Injury
Date Received
April 11, 2012
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S084
Removal / Correction Number
Z-0111-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE DEVICE MET 92% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD EARLY BATTERY DEPLETION DUE TO THE COMPETITOR'S LEFT VENTRICULAR LEAD HAVING HIGH THRESHOLDS SINCE IMPLANT. THE DEVICE WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONSULTA CRT-D IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. D224TRK

Patients

Seq Age Sex Outcome Treatment
1 65 YR Hospitalization| R 6947 IMPLANTABLE TACHY LEAD| 1156T COMPETITOR IMPLANTABLE PACING LEAD| 5594 IMPLANTABLE PACING LEAD