BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE
Report
- Report Number
- 1024879-2026-00859
- Event Type
- Malfunction
- Date Received
- May 26, 2026
- Date of Event
- April 25, 2026
- Report Date
- May 2, 2026
- Manufacturer
- BECTON DICKINSON & CO (FRANKLIN LAKES)
- Product Code
- JKA
- UDI-DI
- 30382903686071
- PMA / PMN Number
- K243207
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
E.1: INITIAL REPORTER FACILITY NAME: (B)(6). THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.4. PMA / 510(K)#: K982541. INVESTIGATION SUMMARY: BD DID NOT RECEIVE ANY SAMPLES OR PHOTOS FOR INVESTIGATION. HOWEVER, 10 RETAINED SAMPLES WERE SUBJECTED TO FUNCTIONAL TESTS. ALL SAMPLES PASSED THE TESTING, AND THERE WAS NO EVIDENCE OF SIDE PIERCED SLEEVES, POOR SLEEVE RECOVERY, SLEEVE FALL OFF, OR SLEEVE LEAKAGE. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS NOT BEEN CONFIRMED FOR THE INDICATED FAILURE MODE: SLEEVE LEAKAGE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE RUBBER SLEEVE OF SIX DEVICES LEAKED BLOOD. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 414095 | BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE | TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION | JKA | BECTON DICKINSON & CO (FRANKLIN LAKES) | 5157570 | 30382903686071 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |