FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE

MDR report key: 25291145 · Received May 26, 2026

Report

Report Number
1024879-2026-00858
Event Type
Malfunction
Date Received
May 26, 2026
Date of Event
April 29, 2026
Report Date
May 2, 2026
Manufacturer
BECTON DICKINSON & CO (FRANKLIN LAKES)
Product Code
JKA
UDI-DI
30382903686071
PMA / PMN Number
K243207
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E.1: INITIAL REPORTER FACILITY NAME: (B)(6). THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.4. PMA / 510(K)#: K982541. INVESTIGATION SUMMARY: BD DID NOT RECEIVE ANY SAMPLES OR PHOTOS FOR INVESTIGATION. HOWEVER, 10 RETAINED SAMPLES WERE SUBJECTED TO FUNCTIONAL TESTS. ALL SAMPLES PASSED THE TESTING, AND THERE WAS NO EVIDENCE OF SIDE-PIERCED SLEEVES, POOR SLEEVE RECOVERY, SLEEVE FALL-OFF, OR SLEEVE LEAKAGE. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS NOT BEEN CONFIRMED FOR THE INDICATED FAILURE MODE: SLEEVE LEAKAGE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE RUBBER SLEEVE OF TWO DEVICES LEAKED BLOOD. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
339890 BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION JKA BECTON DICKINSON & CO (FRANKLIN LAKES) 5161802 30382903686071

Patients

Seq Age Sex Outcome Treatment
1