CLARITY II
Report
- Report Number
- 3021199089-2026-00035
- Event Type
- Injury
- Date Received
- May 26, 2026
- Date of Event
- April 16, 2026
- Report Date
- May 26, 2026
- Manufacturer
- CYNOSURE LUTRONIC TECHNOLOGY CORPORATION
- Product Code
- GEX
- UDI-DI
- 08800291820040
- PMA / PMN Number
- K183566
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THE CYNOSURE LUTRONIC CLINICAL TEAM FOLLOWED UP WITH THE TREATMENT CLINIC FOR ADDITIONAL INFORMATION. IMAGES WERE PROVIDED AND REVIEWED BY CYNOSURE LUTRONIC MEDICAL DIRECTOR. THE CYNOSURE LUTRONIC MEDICAL DIRECTOR RECOGNIZED A FULL THICKNESS BURN ON THE MID CHEEK WITH LOSS OF EPIDERMIS, DERMIS AND SUBCUTANEOUS TISSUES. HE STATED THAT A SCAR IS INEVITABLE, AND IT CANNOT BE PREVENTED. HE ALSO RECOMMENDED THAT THE WOUND NEEDS TO BE CLEANED DAILY AND MANAGED CLOSELY FOR APPROPRIATE CARE AND THE PATIENT TO BE EVALUATED BY A BURN SPECIALIST OR A DERMATOLOGIST. A TRAINED CYNOSURE LUTRONIC FIELD SERVICE ENGINEER PAID A VISIT TO THE TREATMENT FACILITY FOR A DEVICE EVALUATION. DURING THIS TIME THE DEVICE WAS FULLY INSPECTED, AND THERE WAS NO APPARENT DAMAGE TO THE DEVICE BUT THERE WERE A COUPLE PARTS THAT WERE DIRTY BUT THEY WERE CLEANED. THE DEVICE WAS ALSO SUBJECTED TO A FULL DEVICE EVALUATION INCLUDING FULL FUNCTIONALITY TESTING WITH PASSING RESULTS. THE ENERGY LEVELS AND SETTINGS WERE CHECKED AND VERIFIED TO BE WORKING WITHIN ITS SPECIFICATIONS. AT THIS TIME, THE EXACT ROOT CAUSE OF THE EVENT REMAINS UNDETERMINED. WHILE THE TREATMENT PARAMETERS WERE LARGELY WITHIN RECOMMENDED RANGES, FACTORS INCLUDING THE ABSENCE OF TEST SPOTS AND DEVIATION IN HANDPIECE SIZE WERE IDENTIFIED; HOWEVER, A DEFINITIVE CAUSAL RELATIONSHIP CANNOT BE ESTABLISHED. THIS EVENT IS REPORTABLE UNDER FDA MEDICAL DEVICE REPORTING REQUIREMENTS (21 CFR PART 803), AS THE REPORTED INJURY INVOLVED A SERIOUS INJURY REQUIRING MEDICAL INTERVENTION WITH LIKELY PERMANENT IMPAIRMENT DUE TO SCARRING.
IT WAS REPORTED THAT THE PATIENT EXPERIENCED BURN ON THE FACE AFTER UNDERGOING FACIAL TREATMENTS WITH THE CLARITY II DEVICE. THE PATIENT WAS EVALUATED BY A PHYSICIAN, DIAGNOSED WITH A FACIAL BURN, AND PRESCRIBED A TOPICAL ANTIBIOTIC AS PREVENTATIVE THERAPY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 492173 | CLARITY II | CLARITY II | GEX | CYNOSURE LUTRONIC TECHNOLOGY CORPORATION | 08800291820040 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |