FDA Adverse Event
Malfunction
Summary report: N
FREESTYLE LIBRE 2 PLUS
MDR report key: 25280216
·
Received May 26, 2026
Report
- Report Number
- 2954323-2026-94743
- Event Type
- Malfunction
- Date Received
- May 26, 2026
- Date of Event
- April 30, 2026
- Report Date
- May 26, 2026
- Manufacturer
- ABBOTT DIABETES CARE INC
- Product Code
- QBJ
- UDI-DI
- 00357599871000
- PMA / PMN Number
- K223435
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 0
THE PRODUCT HAS BEEN REQUESTED BACK FOR INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF INVESTIGATION ACTIVITIES OR IF ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
Description of Event or Problem · 0
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 190, 104 MG/DL COMPARED TO READINGS OF 270, 225 MG/DL RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 1 DAY. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 126575 | FREESTYLE LIBRE 2 PLUS | FLASH GLUCOSE MONITORING SYSTEM | QBJ | ABBOTT DIABETES CARE INC | 78747-01 | 00357599871000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |