FDA Adverse Event Malfunction Summary report: N

XERF 2.0

MDR report key: 25279158 · Received May 26, 2026

Report

Report Number
1222993-2026-00040
Event Type
Malfunction
Date Received
May 26, 2026
Date of Event
April 27, 2026
Report Date
May 26, 2026
Manufacturer
CYNOSURE LUTRONIC TECHNOLOGY CORPORATION
Product Code
GEI
UDI-DI
08800291821795
Product Problem
Yes
Report Source
Distributor report
Reporter Location
CA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Description of Event or Problem · 0

IT WAS REPORTED THAT PATIENT EXPERIENCED A BURNING SENSATION AND A VISIBLE MARK APPEARED ON PATIENT'S SKIN FOLLOWING A TREATMENT USING XERF. SITE ALSO RELAYED THAT TREATMENT WAS PERFORMED AT LEVEL 7 AFTER 20 PULSES. CYNOSURE LUTRONIC (CL) CLINICAL TEAM CONTACTED THE SITE TO INVESTIGATE THE EVENT. CL CLINICAL DETERMINED THE EVENT TO BE INCONCLUSIVE. PATIENT'S TREATMENT COULD NOT BE COMPLETED WITH 100 PULSES REMAINING IN THE TREATMENT AREA. TREATMENT WAS STOPPED AFTER PATIENT COMPLAINED OF BURNING SENSATION. IMAGES OF THE EFFECTOR TIP WERE PROVIDED. PER CL CLINICAL, "THE EFFECTOR TIP APPEARS TO HAVE A SCORCHED AREA. IT CANNOT BE CONFIRMED IF THE EFFECTOR WAS DEFECTIVE, OR DURING THE CLEANING PROCESS AND REMOVING THE EFFECTOR FROM THE PACKAGE IT WAS DAMAGED." CL CLINICAL REMINDED SITE OF PROPER CARE OF THE EFFECTOR PER LABELING. XERF EFFECTOR TIP HAS NOT YET BEEN RETURNED FOR A DEVICE EVALUATION, BUT A PHOTO WAS PROVIDED DURING INITIAL COMMUNICATION. SURFACE DAMAGE WAS OBSERVED ON THE TIP. ONCE TIP HAS BEEN RETURNED AND EVALUATED, A CYNOSURE FIELD SERVICE ENGINEER (FSE) CAN DETERMINE ROOT CAUSE. THE EVENT IS REPORTABLE PER FDA MEDICAL DEVICE REPORTING TITLE 21 CFR PART 803 SINCE AN OBSERVED MALFUNCTION OCCURRED AND COULD LIKELY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY IF THE MALFUNCTION WERE TO RECUR. WE ARE REPORTING THIS AS AN INITIAL MDR AND WILL SUBMIT A FINAL MDR ONCE INVESTIGATION HAS BEEN COMPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
351200 XERF 2.0 XERF 2.0 GEI CYNOSURE LUTRONIC TECHNOLOGY CORPORATION 08800291821795

Patients

Seq Age Sex Outcome Treatment
1 39 YR Female