BREATHING CIRCUIT SET

MDR report key: 25277198 · Received May 26, 2026

Report

Report Number
25277198
Event Type
Malfunction
Date Received
May 26, 2026
Date of Event
March 7, 2026
Report Date
May 14, 2026
Manufacturer
HAMILTON MEDICAL AG
Product Code
BZO
Product Problem
Yes
Report Source
User Facility report
Reporter Location
CT
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Description of Event or Problem · 0

WHILE ENROUTE TO THE OUTSIDE HOSPITAL, WE WERE SETTING UP THE VENTILATOR CIRCUIT. THE VIRAL FILTERS WERE NOT ATTACHING BETWEEN THE EXHALATION LIMB OF THE CIRCUIT AND THE EXHALATION VALVE WHERE THEY NORMALLY GO. WE NOTICED THAT THE ELBOW OF THE EXPIRATORY LIMB HAD AN INNER LUMEN THAT THE INSPIRATORY LIMB ELBOW DID NOT. WE CHECKED THE OTHER LARGE CIRCUIT IN THE TRUCK, AND IT DID NOT HAVE THIS EITHER. WE WERE ABLE TO ASSEMBLE THE CIRCUIT WITH THE FILTER AT THE PATIENT WYE, SO IT FILTERED BOTH THE INSPIRATORY AND EXPIRATORY LIMBS. WE WERE ALSO ABLE TO SETUP THE AEROGEN BETWEEN THE PATIENT AND FILTER, SO THE VENTILATOR WAS PROTECTED. WITH THIS CONFIGURATION, THE VENT DID PASS ALL PRE-OP CHECKS. IT WORKED AS INTENDED FOR THE PATIENT. AFTER WE COMPLETED THE TRANSFER, WE LOOKED AT A FEW MORE CIRCUITS WITH THE SAME REF (BUT DIFFERENT LOT) AND IDENTIFIED THAT THIS ONE PARTICULAR CIRCUIT HAD THE EXPIRATORY LIMB ELBOW FLIPPED. THE CIRCUIT WAS NOT SAVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
417261 BREATHING CIRCUIT SET SET, TUBING AND SUPPORT, VENTILATOR (W HARNESS) BZO HAMILTON MEDICAL AG UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1