FREESTYLE LIBRE 2 PLUS
Report
- Report Number
- 2954323-2026-90078
- Event Type
- Malfunction
- Date Received
- May 23, 2026
- Date of Event
- April 27, 2026
- Report Date
- May 23, 2026
- Manufacturer
- ABBOTT DIABETES CARE INC
- Product Code
- QBJ
- UDI-DI
- 00357599871000
- PMA / PMN Number
- K223435
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- 003
Narratives
THE PRODUCT HAS BEEN REQUESTED BACK FOR INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF INVESTIGATION ACTIVITIES OR IF ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 113, 91, 53 MG/DL COMPARED TO READINGS OF 113, 137, 267 MG/DL RESPECTIVELY OBTAINED ON A BUILT-IN PRECISIONAND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 9 DAYS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 505498 | FREESTYLE LIBRE 2 PLUS | FLASH GLUCOSE MONITORING SYSTEM | QBJ | ABBOTT DIABETES CARE INC | 78747-01 | 00357599871000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |