BECKMAN COULTER ¿ COULTER ® LH 750 SLIDEMAKER
Report
- Report Number
- 1061932-2012-01183
- Event Type
- Malfunction
- Date Received
- April 11, 2012
- Date of Event
- March 12, 2012
- Report Date
- March 12, 2012
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- GKJ
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4)
ROOT CAUSE FOR THE EVENT IS DUE PINCH TUBING CUT DUE TO WEAR AT THE PINCH POINT. (B)(4).
(B)(4)
THE CUSTOMER REPORTED SMALL AMOUNT OF PINK-TINGED FLUID FROM BECKMAN COULTER LH 750 SLIDEMAKER INSTRUMENT ONTO THE FLOOR AND THE TABLE. THE LEAK WAS NOT CONTAINED WITHIN THE INSTRUMENT AND THE SOURCE OF THE LEAK WAS UNKNOWN. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF A LAB COAT, EYE PROTECTION, AND GLOVES. THE CUSTOMER DID NOT COME IN CONTACT WITH THE FLUID AND DID NOT SEEK MEDICAL ATTENTION. NO EXPOSURE (SPRAYED OR SPLASHED) TO MUCOUS MEMBRANES OR OPEN WOUNDS WERE REPORTED. NO DEATH, INJURY OR CHANGES TO PATIENT TREATMENT OR RESULTS ARE ASSOCIATED WITH THIS EVENT. THE FACILITY DOES HAVE A RISK MANAGEMENT / EXPOSURE CONTROL PLAN IS IN PLACE. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ADDRESS THE EVENT. UPON ARRIVAL THE FSE STATED THE UNIT WAS TURNED OFF AND NO LEAKS WERE FOUND. THE LEAK HAD NOT BEEN CONTAINED WITHIN THE UNIT; THE CUSTOMER HAD CLEANED FLUID FROM THE TABLE. THE DRIP TRAY IN THE UNIT WAS FULL OF FLUID; APPROXIMATELY 8 OZ. A PIECE OF BLACK STRIPE NC TUBING THROUGH PV19 WAS LEAKING DUE TO WEAR AT THE PINCH POINT. THE TUBING SUPPLIED DILUENT/CLENZ/BLEACH UNDER PRESSURE. NO OTHER PARTS WERE LEAKING. THE UNIT WAS PERFORMING WITHIN SPECIFICATIONS AFTER THE SERVICE. THE LINE SUPPLIED DILUENT, CLENZ AND BLEACH; THE SOLUTION ALSO MIXED WITH DRIED SOLUTION FROM PREVIOUS LEAKS. THE FIELD SERVICE ENGINEER (FSE) CLEANED THE UNIT AND REPLACED BLACK STRIPE TUBING CUT AT PINCH VALVE PV19.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BECKMAN COULTER ¿ COULTER ® LH 750 SLIDEMAKER | SPINNER, SLIDE, AUTOMATED | GKJ | BECKMAN COULTER INC. | LH750 SM | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |