FDA Adverse Event Malfunction Summary report: N

ENDOWRIST

MDR report key: 25263471 · Received May 23, 2026

Report

Report Number
2955842-2026-26183
Event Type
Malfunction
Date Received
May 23, 2026
Date of Event
April 8, 2026
Report Date
May 22, 2026
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874112298
PMA / PMN Number
K220023
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT WAS ANALYZED AND FOUND TO HAVE A MELTED BLADE TIP. VISUAL INSPECTION FOUND THE INSTRUMENT BLADE MELTED AT TIP. THE INSTRUMENT WAS PLACED A DRIVEN ON AN IN-HOUSE SYSTEM. THE INSTRUMENT PASSED THE RECOGNITION AN ENGAGEMENT TESTS. THE INSTRUMENT MOVED INTUITIVELY WITH FULL RANGE OF MOTION IN ALL DIRECTIONS. THE GRIPS DID NOT STRUGGLE TO BE OPENED OR CLOSED. THE INSTRUMENT RELEASED ENERGY AND BEGAN ARCING ALMOST IMMEDIATELY. THE COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS. THE PROBABLE ROOT CAUSE OF THE THERMAL DAMAGE IS PRIMARILY ATTRIBUTED TO THE USE CONDITIONS OF MONOPOLAR ENERGY IF THE LOWEST POWER OR EFFECT SETTING POSSIBLE IS NOT USED FOR THE MINIMUM TIME NECESSARY TO ACHIEVE THE DESIRED EFFECT. EXCESSIVE ENERGY USAGE CAN CAUSE MELTING OF THE BLADES.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE SCISSORS TIP MELTED INTO THE 8MM MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT. A BACKUP INSTRUMENT WAS USED FOR THE PROCEDURE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THERE WAS NO ARCING OBSERVED. THERE WAS NO PATIENT INJURY OR HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
551316 ENDOWRIST MONOPOLAR CURVED SCISSORS NAY INTUITIVE SURGICAL, INC 470179-23 K11251031 0405 00886874112298

Patients

Seq Age Sex Outcome Treatment
1 73 YR Male DA VINCI INSTRUMENTS AND ACCESSORIES.