BD PYXIS¿ MEDSTATION¿ ES
Report
- Report Number
- 2016493-2026-31008
- Event Type
- Malfunction
- Date Received
- May 22, 2026
- Date of Event
- April 26, 2026
- Report Date
- May 2, 2026
- Manufacturer
- CAREFUSION 303, INC.
- Product Code
- BRY
- UDI-DI
- 10885403533228
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A REVIEW OF THE COMPLAINT HISTORY FOR SN (B)(6) WAS PERFORMED IN SALESFORCE WHICH DID NOT LOCATE SIMILAR COMPLAINT(S) WITH THE SAME FAILURE MODE FOR THIS SERIAL NUMBER. A REVIEW OF THE DEVICE HISTORY RECORD FOR SN (B)(6) WAS PERFORMED FROM THE DATE OF MANUFACTURE, 20-MAY-2019 AND CONFIRMED THAT THIS DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICING AND THERE WERE NO PRODUCTION FAILURES WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. UPON INVESTIGATION OF THE ACTUAL DEVICE USED IN THIS INCIDENT, IT WAS DETERMINED THAT THE MATRIX DRAWER WAS NOT OPENING. A FIELD SERVICE ENGINEER POWERED DOWN THE STATION AND REMOVED THE BACK PANEL AND FOUND THE U LINK PLUNGER WAS BROKEN. REMOVED THE DRAWER FROM THE CABINET, REPLACED THE PLUNGER, REINSTALLED THE DRAWER INTO THE STATION, CLOSED THE BACK PANEL, AND POWERED THE STATION BACK ON. THE SYSTEM FUNCTIONED AS INTENDED AFTER THE FIELD SERVICE ENGINEER REPAIRED THE DEVICE.
IT WAS REPORTED THAT WHEN USING THE BD PYXIS¿ MEDSTATION¿ ES MBU MAIN DRAWER 1 WOULD NOT OPEN FOR DISPENSING OR RECOVERY. THE CUSTOMER OPENED THE BACK OF THE UNIT AND WAS ABLE TO PUSH THE DRAWER OUT USING THE RED TAB, WITH NO OBSTRUCTIONS OR SCRAPING OBSERVED DURING OPENING. THE CUSTOMER ALSO REBOOTED AND PERFORMED A FULL POWER CYCLE; HOWEVER, NEITHER ACTION RESOLVED THE ISSUE. THE CUSTOMER STATED THAT THERE WAS A DELAY IN PATIENT CARE. THERE WERE NO ADVERSE EVENTS OR INJURIES REPORTED BASED ON THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 236432 | BD PYXIS¿ MEDSTATION¿ ES | AUTOMATED DISPENSING CABINET | BRY | CAREFUSION 303, INC. | 500001002500 | 10885403533228 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |