FDA Adverse Event Injury Summary report: N

HANDY SOLUTIONS

MDR report key: 25250529 · Received May 21, 2026

Report

Report Number
3030539818-2026-00012
Event Type
Injury
Date Received
May 21, 2026
Date of Event
February 27, 2026
Report Date
March 27, 2026
Manufacturer
NINGBO UNICO PRODUCTS CO., LTD 
Product Code
IRT
UDI-DI
00024291256076
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHARMACIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE MANUFACTURER RECEIVED A REPORT STATING THAT THE HEATING PAD OVERHEATED, ALONG WITH PHOTOS OF THE INCIDENT. THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE PRODUCT LABEL WARNED AGAINST USE ON SOFT SURFACE AND CAN'T BE COVERED DURING USE, YET THESE ACTIONS WERE OBSERVED IN THE INCIDENT PHOTOS. THIRD-PARTY TESTING CONFIRMED THE PRODUCT DOES NOT OVERHEAT UNDER CORRECT USE (SEE TEST REPORT: ZX260209-C060401) AND NO DEFECTS IDENTIFIED.

Description of Event or Problem · 0

PATIENT CALLED TO REPORT RECEIVING LETTER ADVISING OF OVERHEATING ISSUE WITH HEATING PAD (OTC#448) PURCHASED FROM PHARMACY OTC DEPARTMENT. PATIENT STATES SHE USED THE HEATING PAD A FEW TIMES SINCE 2025. PATIENT STATES SHE LIVES IN, THEREFORE DOES NOT NEED TO USE HEATING PAD MUCH DUE TO WARM TEMPERATURES. PATIENT STATES SHE NOTICED BURN MARKS ON MATTRESS 3 TO 4 WEEKS AGO. PATIENT STATES SHE DID NOT REPORT IT BECAUSE SHE THOUGHT IT WAS A "FLUKE". PATIENT STATES SHE DECIDED TO REPORT INCIDENT, AFTER RECEIVING LETTER NOTIFYING HER OF AN OVERHEATING DEFECT OF THE PAD EXPLAINED IN THE LETTER FROM. PATIENT REQUESTING REPLACEMENT OF HER MATTRESS. PATIENT DID NOT SUFFER ANY PERMANENT BODILY INJURY. PATIENT STATES SHE STILL HAS THE MATTRESS AND HEATING PAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321105 HANDY SOLUTIONS PAD, HEATING, POWERED IRT NINGBO UNICO PRODUCTS CO., LTD  25607 V1 00024291256076

Patients

Seq Age Sex Outcome Treatment
1