FDA Adverse Event Injury Summary report: N

HANDY SOLUTIONS

MDR report key: 25250525 · Received May 21, 2026

Report

Report Number
3030539818-2026-00013
Event Type
Injury
Date Received
May 21, 2026
Date of Event
March 26, 2026
Report Date
March 26, 2026
Manufacturer
NINGBO UNICO PRODUCTS CO., LTD 
Product Code
IRT
UDI-DI
00024291256076
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHARMACIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE MANUFACTURER RECEIVED A REPORT STATING THAT THE HEATING PAD OVERHEATED, ALONG WITH PHOTOS OF THE INCIDENT. THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE PRODUCT LABEL WARNED AGAINST USE ON SOFT SURFACE AND CAN'T BE COVERED DURING USE, YET THESE ACTIONS WERE OBSERVED IN THE INCIDENT PHOTOS. THIRD-PARTY TESTING CONFIRMED THE PRODUCT DOES NOT OVERHEAT UNDER CORRECT USE (SEE TEST REPORT (B)(4) AND NO DEFECTS IDENTIFIED. THIRD PART TESTING REPORT (B)(4).

Description of Event or Problem · 0

PATIENT USING AN OTC HEATING PAD WRAP FOR SHOULDER AND NECK FROM MANUFACTURER-NAVAJO MANUFACTURING COMPANY INC. ON LOW SETTING AND EXPERIENCED PRODUCT OVERHEATING . PATIENT EXPERIENCED BURNING AND BLISTERING TO NECK AND SHOULDER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
382382 HANDY SOLUTIONS PAD, HEATING, POWERED IRT NINGBO UNICO PRODUCTS CO., LTD  25607 V1 00024291256076

Patients

Seq Age Sex Outcome Treatment
1