FDA Adverse Event Injury Summary report: N

IMPELLA

MDR report key: 25240758 · Received May 21, 2026

Report

Report Number
1220648-2026-08065
Event Type
Injury
Date Received
May 21, 2026
Date of Event
May 12, 2026
Manufacturer
ABIOMED, INC. - 1220648
Product Code
OZD
UDI-DI
00813502012279
PMA / PMN Number
P140003
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ABIOMED INC OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ABIOMED INC, OR ITS EMPLOYEES, CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

AN IMPELLA CP DEVICE WAS INSERTED INTO THE RIGHT FEMORAL ARTERY IN A 70-YEAR-OLD MALE PATIENT PRESENTING IN ACUTE MYOCARDIAL INFARCTION (AMI) AND CARDIOGENIC SHOCK (CGS), IN SCAI E SHOCK, AND ON MULTIPLE INOTROPES AND VASOPRESSORS, PRIOR TO INITIATION OF SUPPORT. WHILE THE PATIENT WAS IN THE INTENSIVE CARE UNIT (ICU), THERE WAS HEMOLYSIS WITH PINK/LIGHT RED URINE. THE IMPELLA WAS WEANED DOWN FROM P-6 TO P-4 AND RECEIVED FLUID BOLUSES TO INCREASE THE CENTRAL VENOUS PRESSURE (CVP). IT IS UNKNOWN IF THE HEMOLYSIS RESOLVED. AFTER ONE DAY ON SUPPORT, THE DEVICE WAS SUCCESSFULLY WEANED. THE POST-PROCEDURE OUTCOME IS SURVIVED AT EXPLANT. WHILE ON SUPPORT, THE DEVICE FUNCTIONED AS INTENDED AT P-4 AT 2.2L/MIN WITH D5W SODIUM BICARBONATE IN THE PURGE SOLUTION. HEMOLYSIS CAN BE ATTRIBUTED TO MALPOSITION OF THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
432087 IMPELLA TEMPORARY NON-ROLLER TYPE LEFT HEART SUPPORT BLOOD PUMP OZD ABIOMED, INC. - 1220648 2027867975 00813502012279

Patients

Seq Age Sex Outcome Treatment
1 70 YR Male Required Intervention