FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 2523566 · Received April 9, 2012

Report

Report Number
2531779-2012-02949
Event Type
Malfunction
Date Received
April 9, 2012
Report Date
March 11, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

LOT # B201761.THE CARTRIDGES HAVE NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4): THE CARTRIDGES PASSED VISUAL INSPECTION; NO DAMAGE OR DEFECTS WERE OBSERVED TO THE LUER LOCK, O-RINGS, PLUNGER, OR CARTRIDGE BODIES. A FILL TEST WAS COMPLETED WITH NO AIR BUBBLES BEING FORMED; THE CARTRIDGE CYCLED NORMALLY AND NO DIFFICULTIES WERE FOUND FILLING THE CARTRIDGE. A LEAK TEST WAS PERFORMED AND NO LEAKS WERE OBSERVED FROM THE LUER CONNECTION, O-RINGS, OR CARTRIDGE BODIES.

Description of Event or Problem · 1

THE PATIENT REPORTED ON (B)(6) THAT SHE HAS USED THREE FAULTY CARTRIDGES AND WANTS TO SEND THEM BACK AND HAVE THEM REPLACED. SHE SAYS SHE IS GETTING AIR BUBBLES WITH THE CARTRIDGES. THE PATIENT IS LEAVING THE VIAL UPRIGHT ON THE COUNTER THEN INSERTING A NEEDLE TO EXPEL AIR THEN FLIPS THE VIAL OVER TO DRAW IN INSULIN. THIS TECHNIQUE IS RECOMMENDED BY THE IFU. HOWEVER THE PATIENT IS NOT LIGHTLY TAPPING THE SYSTEM AND NOT GETTING ALL THE AIR OUT OF THE SYRINGE AS RECOMMENDED BY THE IFU. THE FILLING NEEDLE WAS TIGHTENED SECURELY TO THE CARTRIDGE. THE INSULIN IS A ROOM TEMPERATURE. THE PLUNGER WAS CYCLED AS INSTRUCTED IN THE IFU. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED AS IT WAS ALLEGED THAT THERE WERE AIR BUBBLES IN THE CARTRIDGE THAT COULD NOT BE EXPELLED UPON FILLING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN CARTRIDGE INSULIN PUMP CARTRIDGE LZG ANIMAS CORPORATION ANM IR1200/1250/2020/OTP CART

Patients

Seq Age Sex Outcome Treatment
1 72 YR