FDA Adverse Event Malfunction Summary report: N

CLINICAL CHEMISTRY CALCIUM

MDR report key: 2523438 · Received April 9, 2012

Report

Report Number
1628664-2012-00222
Event Type
Malfunction
Date Received
April 9, 2012
Date of Event
March 23, 2012
Report Date
March 23, 2012
Manufacturer
ABBOTT MANUFACTURING INC
Product Code
CJY
PMA / PMN Number
K981578
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN EVALUATION IS IN PROCESS. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Additional Manufacturer Narrative · 1

THE INSTRUMENT SERVICE HISTORY WAS REVIEWED AND BASED ON THE AVAILABLE INFORMATION, NO OTHER TICKETS WERE FOUND RELATED TO THIS COMPLAINT. A SEARCH FOR SIMILAR COMPLAINTS WAS PERFORMED; ONE OTHER COMPLAINT FOR ERRATIC CALCIUM RESULTS WAS CLOSED WITH NO DEFICIENCY, AND TWO OTHER TICKETS FOR ERRATIC CALCIUM RESULTS WERE CLOSED THROUGH NORMAL TROUBLESHOOTING OR ARE STILL IN-PROCESS. THE ARCHITECT CALCIUM REAGENT PACKAGE INSERT AND ARCHITECT SYSTEM OPERATIONS MANUAL CONTAIN ADEQUATE LABELING FOR SAMPLE PRECAUTIONS AND TROUBLESHOOTING OF ERRATIC RESULTS. BASED UPON THE AVAILABLE DATA AND THE RESULTS OF THIS EVALUATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

THE CUSTOMER STATED A FALSELY ELEVATED CLINICAL CHEMISTRY CALCIUM RESULT OF 3.15 MMOL/L (12.6 MG/DL) WAS GENERATED WHEN REAGENT LOT 06879UN11 WAS IN USE. REPEAT TESTING OF THE SAMPLE GENERATED A RESULT OF 2.36 MMOL/L (9.4 MG/DL), WITH RESULTS OF 2.34 MMOL/L (9.36 MG/DL) GENERATED ON A SECOND ARCHITECT IN THE LAB. THE CUSTOMER WAS ADVISED TO PERFORM A CONTAMINATION TEST WITH CREATININE AND CALCIUM TO DETERMINE IF THERE WAS A PROBLEM. THE CUSTOMER PERFORMED FIVE RUNS OF THE CONTAMINATION TEST AND NO ISSUES WERE IDENTIFIED. THE FALSELY ELEVATED CALCIUM RESULT WAS NOT REPORTED TO THE MEDICAL PROVIDER, THEREFORE, THERE WAS NO IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CLINICAL CHEMISTRY CALCIUM CJY ABBOTT MANUFACTURING INC 06879UN11

Patients

Seq Age Sex Outcome Treatment
1 ARCHITECT C8000, LIST #1G06-01, SN (B)(4)| ARCHITECT C8000, LIST#1G06-01, SN (B)(4)