FDA Adverse Event Malfunction Summary report: N

SMARTDRIVE

MDR report key: 25233950 · Received May 20, 2026

Report

Report Number
3008370857-2026-00339
Event Type
Malfunction
Date Received
May 20, 2026
Date of Event
April 5, 2026
Report Date
May 20, 2026
Manufacturer
MAX MOBILITY LLC
Product Code
ITI
UDI-DI
00861896000310
PMA / PMN Number
K151199
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

PERMOBIL AUSTRALIA RECEIVED A REPORT INDICATING THE END-USER WAS IN A STORE AND HAD STOPPED THE SMARTDRIVE DEVICE USING THE SWITCH CONTROL BUTTON W/MONO JACK, THEN SUDDENLY THE SMARTDRIVE REPORTEDLY ENGAGED A DRIVE COMMAND WITHOUT ANY PHYSICAL MANIPULATION OF THE SWITCH AND WAS REPORTEDLY UNABLE TO STOP WHEN PRESSING THE SWITCH. THIS FORCED THE END-USER TO PULL THE SWITCH CONTROL BY THE PLUG END IN ORDER TO STOP. NO INJURIES WERE REPORTED TO HAVE BEEN SUSTAINED AS A RESULT OF THE EVENT. SERVICE PROVIDER REPORTED TO PERMOBIL THAT THEY WERE INITIALLY UNABLE TO REPLICATE A FAILURE, THEN LATER REPORTED TO HAVE REPLACED THE CABLE UNDER WARRANTY. PERMOBIL AUSTRALIA HAS REQUESTED THE SUSPECT COMPONENT BE RETURNED FOR EVALUATION, AND AT THIS TIME PERMOBIL HAS NOT RECEIVED ANY PRODUCT TO EVALUATE AND IS UNABLE TO REACH A DETERMINATION AS TO POSSIBLE ROOT CAUSE WITHOUT SPECULATION. IF ANY NEW INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

REPORT OF SWITCH CONTROL W/MONO JACK AND BUDDY BUTTON HAVING BEEN PRESSED TO STOP, THEN UNEXPECTEDLY ENGAGED THE SMARTDRIVE MOTOR TO WHERE THE END-USER REPORTEDLY COULD NOT STOP WITH A PRESS OF THE CONTROLLER BUTTONS. NO INJURY WAS REPORTED TO HAVE OCCURRED AS A RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
560320 SMARTDRIVE SMARTDRIVE ITI MAX MOBILITY LLC MX2+ 00861896000310

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown