FDA Adverse Event Malfunction Summary report: N

ALINITY I STAT HIGH SENSITIVITY TROPONIN-I

MDR report key: 25232520 · Received May 20, 2026

Report

Report Number
3005094123-2026-00239
Event Type
Malfunction
Date Received
May 20, 2026
Date of Event
April 25, 2026
Report Date
May 20, 2026
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
MMI
UDI-DI
00380740194062
PMA / PMN Number
K202525
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED A FALSE POSITIVE ALINITY I HIGH SENSITIVE TROPONIN RESULT GENERATED ON AN ALINITY I PROCESSING MODULE FOR A 39-YEAR-OLD MALE PATIENT. SID (B)(6) INITIAL RESULT = 87.75 NG/L. A NEW SAMPLE, SID (B)(6), WAS DRAWN GENERATING A RESULT <2.70 NG/L. BOTH SAMPLES WERE REPEATED AND GENERATED A RESULT OF <2.70 NG/L. THE CUSTOMER¿S REFERENCE RANGE IS 0-53 NG/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
560109 ALINITY I STAT HIGH SENSITIVITY TROPONIN-I IMMUNOASSAY METHOD, TROPONIN SUBUNIT MMI ABBOTT IRELAND DIAGNOSTICS DIVISION 85393UD00 00380740194062

Patients

Seq Age Sex Outcome Treatment
1 39 YR Male ALNTY I PROCESSING MODU, 03R65-01, AI25780.