FDA Adverse Event Malfunction Summary report: N

TANGO OPTIMO

MDR report key: 2523087 · Received June 25, 2010

Report

Report Number
9610824-2010-00003
Event Type
Malfunction
Date Received
June 25, 2010
Date of Event
May 24, 2010
Report Date
June 22, 2010
Manufacturer
BIO-RAD MED DIAGNOSTICS GMBH
Product Code
KSZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BIO-RAD LABORATORIES INC ACQUIRED THE BIOTEST MEDICAL DIAGNOSTICS (B)(4) ON JANUARY 6, 2010. THE COMPANY NAME WAS CHANGED TO BIO-RAD MEDICAL DIAGNOSTICS ON MARCH 23, 2010.

Description of Event or Problem · 1

ON (B)(6) 2010 AN A NEG PT WAS DETERMINED AS O NEG. DISCREPANCY WAS DETECTED AS RESULT WAS TRANSFERRED TO HOST AND COMPARED WITH HISTORIC GROUP. REPEAT ON SAME SAMPLE ON SAME SYSTEM GAVE A NEG RESULT. THE WRONG RESULT WAS DETERMINED BY TANGO OPTIMO IN ABD CONFIRMATION ASSAY. THE CAUSE IS NOT KNOWN YET. THE CONCERNED SAMPLES WERE FLAGGED BY THE SYSTEM. THE TEMPERATURE IN THE LAB WAS HIGH, BUT NO OTHER SAMPLE SHOWED A WRONG RESULT. ALL REAGENTS USED WITH THE PT HAD BEEN TRANSFERRED TO AN OTHER SYSTEM AND PASSED QC WITHOUT ANY PROBLEM. THE REAGENTS HAD BEEN USED UP SO THEY COULDN'T BE SHIPPED FOR RETESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TANGO OPTIMO AUTOMATED BLOOD GROUP ANALYZER KSZ BIO-RAD MED DIAGNOSTICS GMBH

Patients

Seq Age Sex Outcome Treatment
1