FDA Adverse Event
Malfunction
Summary report: N
ALINITY I TOTAL B-HCG REAGENT KIT
MDR report key: 25228816
·
Received May 20, 2026
Report
- Report Number
- 3005094123-2026-00238
- Event Type
- Malfunction
- Date Received
- May 20, 2026
- Date of Event
- May 10, 2026
- Report Date
- May 20, 2026
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- DHA
- UDI-DI
- 00380740163334
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
AN EVALUATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 7P51-77 HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 7P51-21/-31, AND A 510K/PMA/BLA NUMBER OF K170317. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.
Description of Event or Problem · 0
THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I TOTAL. -HCG RESULT GENERATED FOR A 31-YEAR-OLD FEMALE PATIENT SAMPLE. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE <5.00 MIU/ML IS NEGATIVE): SAMPLE ID (B)(6). INITIAL RESULT = 27.5 MIU/ML, REPEAT RESULTS = <1.2 MIU/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 67374 | ALINITY I TOTAL B-HCG REAGENT KIT | SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN | DHA | ABBOTT IRELAND DIAGNOSTICS DIVISION | 80026UD01 | 00380740163334 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |