JENAVALVE TECHNOLOGY, INC.
Report
- Report Number
- 3012148004-2026-70002
- Event Type
- Injury
- Date Received
- May 19, 2026
- Date of Event
- April 21, 2026
- Report Date
- May 19, 2026
- Manufacturer
- JENAVALVE TECHNOLOGY, INC.
- Product Code
- NPT
- UDI-DI
- 00850000898110
- PMA / PMN Number
- P250024
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
SOME OF THE PATIENT IDENTIFYING INFORMATION SUCH AS AGE, SEX, AND DATE OF BIRTH IS NOT RECEIVED BY JENAVALVE TECHNOLOGY, INC PER DSGVO REGULATIONS.
A PATIENT-RELATED ADVERSE EVENT OCCURRED ON APRIL 21, 2026, DURING IMPLANTATION OF A TRILOGY-THV-M DEVICE IN (B)(6). DURING POSITIONING OF THE VALVE, THE PATIENT EXPERIENCED HEMODYNAMIC DECOMPENSATION. IN RESPONSE, A FEMORAL-FEMORAL BYPASS WAS INITIATED, RESULTING IN STABILIZATION OF THE PATIENT'S CONDITION. ALTHOUGH FEMORAL-FEMORAL BYPASS ACCESS WAS ESTABLISHED, THE BYPASS MACHINE WAS NOT ACTIVATED; IT REMAINED ON STANDBY AS A PRECAUTIONARY MEASURE. FULL CARDIOPULMONARY BYPASS SUPPORT WAS NOT REQUIRED. PATIENT AGE, SEX, WEIGHT AND RACE WERE NOT AVAILABLE AT THE TIME OF REPORTING. FOLLOWING HEMODYNAMIC RECOVERY, THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE PATIENT REMAINED STABLE POST-PROCEDURE AND WAS WEANED WITHOUT DIFFICULTY. THE TRILOGY TRANSCATHETER HEART VALVE DEVICE IS WELL-POSITIONED AND FUNCTIONING AS INTENDED. ON (B)(6) 2026, THE IMPLANTING PHYSICIAN REPORTED THAT THE PATIENT'S OVERALL RECOVERY HAS BEEN SATISFACTORY AND IS ACTIVELY RECOVERING. THE PHYSICIAN ALSO NOTED THAT THE AORTIC INSUFFICIENCY (AI) MAY HAVE BEEN EXACERBATED BY MANIPULATION OF THE WIRE AND/OR DEVICE OR SHEATH, POTENTIALLY CONTRIBUTING TO THE ACUTE DECOMPENSATION. HOWEVER, THE EXACT MECHANISM REMAINS UNCERTAIN, AND NO DEFINITIVE PREVENTIVE STRATEGY HAS BEEN IDENTIFIED. A POSSIBLE ALTERNATIVE APPROACH, SUCH AS INITIATING POSITIONING AT A HIGHER LEVEL, IS CONSIDERED SPECULATIVE. THE DEVICE WAS NOT RETURNED AS IT WAS DISPOSED PRIOR TO REQUEST. JENAVALVE TECHNOLOGY CONDUCTED A FULL DHR REVIEW WHICH SHOWED THAT ALL COMPONENTS OF THE SYSTEM PASSED ALL INSPECTION. BASED ON THE PROVIDED INFORMATION, ALTHOUGH THIS EVENT IS ATTRIBUTED TO THE PROCEDURE, IT IS NOT SUSPECTED THAT A DEVICE DEFICIENCY OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 38616 | JENAVALVE TECHNOLOGY, INC. | N/A | NPT | JENAVALVE TECHNOLOGY, INC. | TRILOGY-THV-M | 00850000898110 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention |