FDA Adverse Event Malfunction Summary report: N

IVENIX INFUSION SYSTEM

MDR report key: 25220129 · Received May 19, 2026

Report

Report Number
3014732157-2026-02192
Event Type
Malfunction
Date Received
May 19, 2026
Date of Event
April 29, 2026
Report Date
May 26, 2026
Manufacturer
FRESENIUS KABI USA LLC
Product Code
FRN
UDI-DI
00811505030320
PMA / PMN Number
K183311
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Description of Event or Problem · 0

THE FOLLOWING HAS BEEN REPORTED: BIOMED REPORTED: "BLUE SCREEN PUMP FAILURE" OBF PUMP FOR SSM. PUMP NEVER USED ON PT. A PRELIMINARY REVIEW IDENTIFIED THE FOLLOWING ISSUE: PNEUMATIC VALVE ACTUATION FAILURE (VALVE 6) AN ACTIVE INFUSION WAS NOT STOPPED. REPORTING AS A CONSERVATIVE MEASURE. NO ADVERSE EFFECTS WERE REPORTED. ADDITIONAL INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
616825 IVENIX INFUSION SYSTEM LARGE VOLUME PUMP (LVP) FRN FRESENIUS KABI USA LLC LVP-0004 00811505030320

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown