FDA Adverse Event Malfunction Summary report: N

ATTEST STEAM CHEMICAL INTEGRATOR

MDR report key: 25219865 · Received May 19, 2026

Report

Report Number
MW5188423
Event Type
Malfunction
Date Received
May 19, 2026
Date of Event
May 14, 2026
Report Date
May 14, 2026
Manufacturer
3M / SOLVENTUM US LLC.
Product Code
JOJ
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

OUR FACILITY HAS IDENTIFIED A RECURRING ISSUE WITH THE 3M ATTEST TYPE 5 STEAM CHEMICAL INTEGRATORS. OVER THE PAST SEVERAL MONTHS, MULTIPLE INTEGRATORS HAVE SHOWN VISIBLE DETERIORATION OF THE GREEN AND RED "ACCEPT" INDICATOR LAYER. THE DEFECTS INCLUDE CRACKING, PEELING, BUBBLING, FLAKING, AND CHIPPING OF THE COATING, WHICH IN SOME CASES EXPOSES THE WHITE BACKING UNDERNEATH. TO DATE, THE ISSUE HAS PRIMARILY BEEN OBSERVED IN PEEL-PACKED ITEMS, THOUGH WE ARE MONITORING WRAPPED SETS AND RIGID CONTAINERS AS A PRECAUTION. WHEN AFFECTED INTEGRATORS ARE OPENED, FLAKES FROM THE INDICATOR LAYER MAY TRANSFER ONTO SURGICAL INSTRUMENTS WITHIN THE PACK. WE HAVE REPORTED THE ISSUE TO SOLVENTUM/3M. THE MANUFACTURER CONFIRMED THEY HAVE RECEIVED SIMILAR REPORTS AND BELIEVE THE PROBLEM MAY BE RELATED TO SPECIFIC STERILIZATION CONDITIONS OR A SUPPLIER-RELATED MATERIAL DEFECT. THEY ADVISED THAT NEWER PRODUCTION LOTS DO NOT SHOW THE ISSUE. AT OUR FACILITY, ANY DEFECTIVE INTEGRATORS ARE IMMEDIATELY REMOVED FROM USE. ALL ITEMS FROM MATCHING LOT NUMBERS ARE QUARANTINED AND RETURNED TO THE MANUFACTURER. INSTRUMENTATION THAT MAY HAVE BEEN EXPOSED TO FLAKES IS REPROCESSED PER MANUFACTURER INSTRUCTIONS. ALTHOUGH 3M HAS STATED THAT STERILIZATION MONITORING ACCURACY IS NOT AFFECTED, OUR STERILE PROCESSING SERVICE CONSIDERS THESE INTEGRATORS NON-USABLE DUE TO THE RISK OF PARTICULATE TRANSFER ONTO INSTRUMENTS. THERE HAVE BEEN NO KNOWN PATIENT SAFETY IMPACTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
402872 ATTEST STEAM CHEMICAL INTEGRATOR INDICATOR, PHYSICAL/CHEMICAL STERILIZATION PROCESS JOJ 3M / SOLVENTUM US LLC. 1243 CC032029

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other