FREESTYLE LIBRE 2
Report
- Report Number
- 2954323-2026-82489
- Event Type
- Malfunction
- Date Received
- May 19, 2026
- Date of Event
- April 21, 2026
- Report Date
- May 19, 2026
- Manufacturer
- ABBOTT DIABETES CARE INC
- Product Code
- QLG
- UDI-DI
- 05021791000876
- PMA / PMN Number
- K193371
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- 003
Narratives
THE PRODUCT HAS BEEN REQUESTED BACK FOR INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF INVESTIGATION ACTIVITIES OR IF ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 54, 54, 54, 54 MG/DL COMPARED TO READINGS OF 142, 154, 88, 171 MG/DL RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 1 DAY. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 523367 | FREESTYLE LIBRE 2 | FLASH GLUCOSE MONITORING SYSTEM | QLG | ABBOTT DIABETES CARE INC | 71992-01 | 05021791000876 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |