FDA Adverse Event Malfunction Summary report: N

EPILME

MDR report key: 25218329 · Received May 19, 2026

Report

Report Number
3015779311-2026-00002
Event Type
Malfunction
Date Received
May 19, 2026
Date of Event
April 13, 2026
Report Date
May 19, 2026
Manufacturer
BERGER AND KRAFT
Product Code
GEX
UDI-DI
00850017374218
PMA / PMN Number
K213261
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

ACCORDING TO THE REPORT SUBMITTED BY THE DISTRIBUTOR, THE HAND PIECE STARTED FIRING BY ITSELF WHILE IN THE HOLDER AND THE TOUCH SCREEN DISPLAY WOULD NOT RESPOND TO TOUCH. THE SAPPHIRE TIP IS FROSTED OVER AND THE FRONT COVER IS MELTED. THE POWER METER INSIDE THE HAND PIECE HOLDER IS DAMAGED. ADVISED THE CLINIC TO NOT USE THE DEVICE AND RETURN FOR INVESTIGATION BY THE MANUFACTURER. THIS IS AN ISOLATED OCCURRENCE AND NO OTHER COMPLAINTS OF THIS NATURE HAVE BEEN REPORTED. NO INJURIES WERE ASSOCIATED WITH THIS EVENT. TECHNICAL INVESTIGATION DETERMINED THAT THE INCIDENT WAS CAUSED BY A MECHANICAL MALFUNCTION OF THE MICRO-SWITCH USED IN THE APPLICATOR TRIGGER ASSEMBLY. THE MICRO-SWITCH REMAINED IN AN ACTIVATED STATE AFTER THE OPERATOR RELEASED THE HAND TRIGGER, RESULTING IN CONTINUOUS TRIGGER SIGNAL DETECTION BY THE SYSTEM. AS A CONSEQUENCE, THE APPLICATOR CONTINUED EMITTING PULSES UNINTENTIONALLY WHILE POSITIONED IN THE DOCKING AREA. INSPECTION OF THE APPLICATOR INDICATED THAT THE MICRO-SWITCH IN THE TRIGGER MECHANISM HAD LIKELY BECOME MECHANICALLY BLOCKED OR FAILED INTERNALLY, PREVENTING PROPER RETURN OF THE MICRO-SWITCH TO ITS RELEASED POSITION. THE ISSUE WAS REPRODUCED DURING FUNCTIONAL TESTING AND CONFIRMED DURING APPLICATOR INSPECTION. THE MAIN DEVICE IS FUNCTIONING AS INTENDED. THE ISSUE IS ISOLATED TO THE HANDPIECE ONLY. A CAPA INVESTIGATION HAS BEEN OPENED TO IDENTIFY ROOT CAUSE AND HOW TO KEEP THIS EVENT FROM REOCCURRING IN THE FUTURE. THIS REPORT WAS SUBMITTED BY OUR IMPORTER UNDER REPORT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
597681 EPILME ALIYA GEX BERGER AND KRAFT APE 00850017374218

Patients

Seq Age Sex Outcome Treatment
1