FDA Adverse Event Malfunction Summary report: N

IPC® HANDPIECE ATTACHMENT - INDIGO¿ DRILL

MDR report key: 25217083 · Received May 19, 2026

Report

Report Number
1625507-2026-01141
Event Type
Malfunction
Date Received
May 19, 2026
Date of Event
April 27, 2026
Report Date
May 19, 2026
Manufacturer
MDT POWERED SURGICAL SOLUTIONS
Product Code
HBE
UDI-DI
00613994933706
PMA / PMN Number
K081475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS CONFIRMED REPORTED ISSUE OF WOBBLING. THE DISTAL BEARINGS WERE FOUND BROKEN AND THE LASER MARKINGS GOT FADED. CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 1845000, SERIAL/LOT #: (B)(6). H3: DEVICE ANALYSIS OF PRODUCT ID: 1845000, SERIAL/LOT #: (B)(6) DID NOT IDENTIFY ANY FAULT. H6: CODES OF IMDRF ANNEX C: C19, ANNEX D: D14 AND ANNEX G: G04063 ARE APPLICABLE FOR PRODUCT ID: 1845000, SERIAL/LOT #: (B)(6). MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE STRAIGHT HANDPIECE HAD A WORN BEARING, CAUSING THE BURR TO WOBBLE DURING USE. THERE WAS NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
293723 IPC® HANDPIECE ATTACHMENT - INDIGO¿ DRILL DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE HBE MDT POWERED SURGICAL SOLUTIONS 1845010 00613994933706

Patients

Seq Age Sex Outcome Treatment
1