CD HORIZON® SOLERA® VOYAGER¿ SPINAL SYSTEM
Report
- Report Number
- 3003120897-2026-00905
- Event Type
- Malfunction
- Date Received
- May 18, 2026
- Date of Event
- February 25, 2026
- Report Date
- May 18, 2026
- Manufacturer
- MEDTRONIC SOFAMOR DANEK USA, INC.
- Product Code
- HWE
- UDI-DI
- 00643169961623
- PMA / PMN Number
- K170679
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H3: PRODUCT ANALYSIS: PRODUCT ID# 6550017 LOT# K24L1142 VISUAL AND OPTICAL EXAMINATION REVEALED THE TIP OF THE INSTRUMENT IS BENT FROM WHAT APPEARS TO BE OVERLOAD. THIS IS CONSISTENT WITH OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, HEALTHCARE PROVIDER) REGARDING A DEVICE USED FOR SPINAL THERAPY. IT WAS REPORTED THAT THE DEVICE WAS DAMAGED. DEFORMATION OF THE TAB EXTENDER CAP CONNECTION REGION OF THE TAB EXTENDER WAS CONFIRMED. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 292345 | CD HORIZON® SOLERA® VOYAGER¿ SPINAL SYSTEM | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED M | HWE | MEDTRONIC SOFAMOR DANEK USA, INC. | 6550017 | K24L1142 | 00643169961623 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |