FDA Adverse Event Injury Summary report: N

ITST FIXATION THREADED GUIDE PIN

MDR report key: 2519926 · Received April 4, 2012

Report

Report Number
1822565-2012-00685
Event Type
Injury
Date Received
April 4, 2012
Date of Event
February 25, 2012
Report Date
March 5, 2012
Manufacturer
ZIMMER, INC.
Product Code
LXH
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVAL SUMMARY: THERE ARE SCRATCHES ON THE SHAFT OF THE PIN AS WELL AS EXTENSIVE DAMAGE ON THE THREADS. DIMENSIONS TAKEN ARE WITHIN SPEC. THE AGE OF THE PIN IS UNK AND THE USAGE HISTORY IS UNK. DETAILS OF SURGICAL TECHNIQUE USED ARE UNK. THERE IS INSUFFICIENT INFO PROVIDED TO DETERMINE THE ROOT CAUSE OF THE REPORTED INCIDENT. THIS PRODUCT WILL BE MONITORED FOR ANY ADVERSE COMPLAINT TRENDS. AS RETURNED, THE TIP IS FRACTURED AT THE THREADED PORTION. THERE ARE SCRATCHES ON THE SHAFT OF THE PIN AS WELL AS EXTENSIVE DAMAGE ON THE THREADS. REVIEW OF THE DEVICE HISTORY RECORDS WAS NOT POSSIBLE AS THE PRODUCT AND/OR LOT NUMBERS REQUIRED FOR RETRIEVAL WERE UNAVAILABLE.

Description of Event or Problem · 1

IT IS REPORTED THAT THE TIP OF THE GUIDE WIRE BROKE OFF AND WENT INTO THE PT DURING AN ITST CASE. ANOTHER INCISION WAS MADE IN ORDER TO RETRIEVE THE BROKEN TIP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ITST FIXATION THREADED GUIDE PIN LXH ZIMMER, INC.

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention