ITST FIXATION THREADED GUIDE PIN
Report
- Report Number
- 1822565-2012-00685
- Event Type
- Injury
- Date Received
- April 4, 2012
- Date of Event
- February 25, 2012
- Report Date
- March 5, 2012
- Manufacturer
- ZIMMER, INC.
- Product Code
- LXH
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
EVAL SUMMARY: THERE ARE SCRATCHES ON THE SHAFT OF THE PIN AS WELL AS EXTENSIVE DAMAGE ON THE THREADS. DIMENSIONS TAKEN ARE WITHIN SPEC. THE AGE OF THE PIN IS UNK AND THE USAGE HISTORY IS UNK. DETAILS OF SURGICAL TECHNIQUE USED ARE UNK. THERE IS INSUFFICIENT INFO PROVIDED TO DETERMINE THE ROOT CAUSE OF THE REPORTED INCIDENT. THIS PRODUCT WILL BE MONITORED FOR ANY ADVERSE COMPLAINT TRENDS. AS RETURNED, THE TIP IS FRACTURED AT THE THREADED PORTION. THERE ARE SCRATCHES ON THE SHAFT OF THE PIN AS WELL AS EXTENSIVE DAMAGE ON THE THREADS. REVIEW OF THE DEVICE HISTORY RECORDS WAS NOT POSSIBLE AS THE PRODUCT AND/OR LOT NUMBERS REQUIRED FOR RETRIEVAL WERE UNAVAILABLE.
IT IS REPORTED THAT THE TIP OF THE GUIDE WIRE BROKE OFF AND WENT INTO THE PT DURING AN ITST CASE. ANOTHER INCISION WAS MADE IN ORDER TO RETRIEVE THE BROKEN TIP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ITST FIXATION THREADED GUIDE PIN | LXH | ZIMMER, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |