FDA Adverse Event Injury Summary report: N

DISTAL RADIOULNAR JOINT ASSEMBLY

MDR report key: 2518311 · Received March 16, 2009

Report

Report Number
3004521401-2009-00002
Event Type
Injury
Date Received
March 16, 2009
Date of Event
February 17, 2009
Report Date
March 13, 2009
Manufacturer
APTIS MEDICAL, LLC.
Product Code
KXE
PMA / PMN Number
K040497
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AFTER RECEIVING THE SCREWS, THEY WERE VISUALLY INSPECTED FOR DEFECTS INCLUDING STRIPPED THREADS AND/OR STRIPPED HEADS. BOTH SCREWS WERE UNDAMAGED AND FUNCTIONED AS DESIGNED WHEN APPLIED TO TEST PIECES. FACTORS POSSIBLY ALLOWING THE LOOSENING OF THE SCREW INCLUDE, ANATOMY INTERPOSED BETWEEN THE MATING SURFACES OF THE DEVICE DURING ASSEMBLY. INSUFFICIENT TIGHTENED OF THE SCREWS DURING INITIAL ASSEMBLY. STRESS EXCEEDING RECOMMENDATIONS DURING USE OF THE DEVICE. THE PT IS AN AVID GOLFER AND WAS INFORMED PRIOR TO INITIAL IMPLANTATION OF THE DEVICE THAT CARE SHOULD BE TAKEN TO LIMIT STRESSES ON THE DEVICE. THE PT ADMITTED TO THE SURGEON, GOLFING ON A FREQUENT BASIS AND HIGH FORCE GROUND STRIKES DURING PLAY. THE KNOWN INCIDENCE OF OCCURRENCE AT THIS TIME IS LESS THE 1% OF IMPLANTED DEVICES, THE POSSIBLE FACTORS WILL BE DISCUSSED WITH USING PHYSICIANS AND CONTINUED MONITORING WILL PROCEED.

Description of Event or Problem · 1

TWO SMALL SCREWS WHICH COMPRISE PARTS OF A DISTAL RADIOULNAR JOINT PROSTHESIS LOOSENED FROM THEIR ASSEMBLED POSITION INSIDE THE DISTAL FOREARM OF A PT. THIS REQUIRED A RE-OPERATION TO REASSEMBLE THE DEVICE. THE SCREWS WERE INSPECTED BY THE SURGEON DURING THE PROCEDURE AND FOUND TO BE FUNCTIONING AS INTENDED. TO ALLOW FURTHER INSPECTION OF THE SCREWS, THEY WERE REPLACED AND THE DEVICE WAS REASSEMBLED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DISTAL RADIOULNAR JOINT ASSEMBLY KXE APTIS MEDICAL, LLC. DRUJ-P30 K08520

Patients

Seq Age Sex Outcome Treatment
1 60 YR Required Intervention