FDA Adverse Event Malfunction Summary report: N

BD PHOENIX¿ NMIC-306

MDR report key: 25176793 · Received May 14, 2026

Report

Report Number
1119779-2026-00779
Event Type
Malfunction
Date Received
May 14, 2026
Date of Event
April 15, 2026
Report Date
April 24, 2026
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
LON
UDI-DI
30382904492923
PMA / PMN Number
SEE H11
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA / 510(K)#: K020322, K022129, K023444, K023634, K023858, K024153, 031530, K031699, K031912, K032299, K032567, K032655, K032675, K033362, K033458, K033558, K033560, K041384, K042932, K052269, K060214, 060217, K060257, K060444, K060447, K061327, K061355, K061867, K062207, K062944, K063301, AND K06348. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING THE BD PHOENIX¿ NMIC-306 A PATIENT ISOLATE HAD A LOW MIC (FALSE SUSCEPTIBLE) RESULT FOR THE DRUG CEFTAZIDIME/AVIBACTAM BUT WHEN REPEATED THE RESULT WAS RESISTANT. THE USER STATED THE ISOLATE WAS PCR-POSITIVE RESULT FOR NEW DELHI METALLO-BETA-LACTAMASE (NDM). THE USER VERIFIED THE FINAL RESULT USING E- TEST. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
435626 BD PHOENIX¿ NMIC-306 SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION LON BECTON DICKINSON & CO. (SPARKS) 6034519 30382904492923

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown