ALINITY I CA 19-9XR REAGENT KIT
Report
- Report Number
- 3002809144-2026-00123
- Event Type
- Malfunction
- Date Received
- May 14, 2026
- Date of Event
- March 24, 2026
- Report Date
- May 14, 2026
- Manufacturer
- ABBOTT GMBH
- Product Code
- NIG
- UDI-DI
- 00380740130527
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. ALL AVAILABLE PATIENT INFORMATION WAS PROVIDED, NO ADDITIONAL PATIENT INFORMATION WAS AVAILABLE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 08P32-20 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 08P32-21, WITH 510K/PMA/BLA NUMBER K052000.
THE CUSTOMER OBSERVED IMPRECISE RESULTS FOR THE ALINITY CA 19-9XR FOR 3 PATIENTS. CLINICAL HISTORY IS UNKNOWN FOR THE PATIENTS. THE FOLLOWING DATA WAS PROVIDED: 73-YEAR-OLD MALE MULTIPLE SAMPLES. CUSTOMER IS QUESTIONING THE RESULT FROM (B)(6) 2026: SID (B)(6), PROCESSED ON (B)(6) 2026 = 1034.27 U/ML. SID (B)(6), PROCESSED ON (B)(6) 2026 = 270.59 U/ML. SID (B)(6), PROCESSED ON (B)(6) 2026 = 236.94 U/ML. HISTORICAL RESULTS: SID (B)(6), PROCESSED ON (B)(6) 2026 = 169.39 U/ML. SID (B)(6), PROCESSED ON (B)(6) 2026 = 180.56 U/ML. SID (B)(6), PROCESSED ON (B)(6) 2026 = 160.12 U/ML. 25-YEAR-OLD FEMALE SID (B)(6), INITIAL RESULT = 40.82 REPEAT RESULT = 74.56 U/ML, 63-YEAR-OLD FEMALE SID (B)(6), INITIAL RESULT = 57.31 REPEAT RESULT = 85.83 U/ML NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 575081 | ALINITY I CA 19-9XR REAGENT KIT | SYSTEM, TEST, CARBOHYDRATE AG (CA19-9), FOR MONIT AND MGMT OF PANCREATIC CANCER | NIG | ABBOTT GMBH | 79769FP00 | 00380740130527 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |