COBAS E 801 ANALYTICAL UNIT
Report
- Report Number
- 1823260-2026-01911
- Event Type
- Malfunction
- Date Received
- May 14, 2026
- Date of Event
- March 30, 2026
- Report Date
- May 14, 2026
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K162606
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE ELECSYS TROPONIN T HS REAGENT LOT NUMBER IS 88252001 (EXPIRATION DATE: 31-JAN-2027).
THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS TROPONIN T HS RESULTS FOR 8 PATIENT SAMPLES TESTED ON THE COBAS E 801 ANALYTICAL UNIT. PATIENT SAMPLE 1 ON (B)(6) 2026: THE INITIAL RESULT WAS 33.3 NG/L. THE FIRST REPEAT RESULT WAS 27.8 NG/L. THE SECOND REPEAT RESULT WAS 27.3 NG/L. PATIENT SAMPLE 2 ON (B)(6) 2026: THE INITIAL RESULT WAS 9.75 NG/L. THE FIRST REPEAT RESULT WAS 5.12 NG/L. THE SECOND REPEAT RESULT WAS 4.74 NG/L. PATIENT SAMPLE 3 ON (B)(6) 2026: THE INITIAL RESULT WAS 8.90 NG/L. THE REPEAT RESULT WAS 5.26 NG/L. PATIENT SAMPLE 4 ON (B)(6) 2026: THE INITIAL RESULT WAS 29.4 NG/L. THE FIRST REPEAT RESULT WAS 25.9 NG/L. THE SECOND REPEAT RESULT WAS 22.0 NG/L. PATIENT SAMPLE 5 ON (B)(6) 2026: THE INITIAL RESULT WAS 41.9 NG/L. THE FIRST REPEAT RESULT WAS 12.0 NG/L. THE SECOND REPEAT RESULT WAS 7.89 NG/L. PATIENT SAMPLE 6 ON (B)(6) 2026: THE INITIAL RESULT WAS 12.9 NG/L. THE REPEAT RESULT WAS 9.79 NG/L. PATIENT SAMPLE 7 ON (B)(6) 2026: THE INITIAL RESULT WAS 6.38 NG/L. THE FIRST REPEAT RESULT WAS 3.00 NG/L. PATIENT SAMPLE 8 ON (B)(6) 2026: THE INITIAL RESULT WAS 17.5 NG/L. THE FIRST REPEAT RESULT WAS 13.0 NG/L. THE SECOND REPEAT RESULT WAS 11.9 NG/L.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 627558 | COBAS E 801 ANALYTICAL UNIT | IMMUNOCHEMISTRY ANALYZER | JJE | ROCHE DIAGNOSTICS | E801 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |