VITROS 5600 INTEGRATED SYSTEM
Report
- Report Number
- 1319681-2012-00054
- Event Type
- Malfunction
- Date Received
- April 4, 2012
- Date of Event
- March 5, 2012
- Report Date
- April 4, 2012
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Product Code
- DLZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION DETERMINED THAT A HIGHER THAN EXPECTED VITROS PHBR QUALITY CONTROL RESULT WAS OBTAINED POST-CALIBRATION OF VITROS PHBR LOT 2537-0058-8452 ON A VITROS 5600 INTEGRATED SYSTEM. THE CUSTOMER RECALIBRATED THE SAME LOT OF VITROS PHBR SLIDES USING ALTERNATE VIALS OF THE CALIBRATOR FLUIDS AS WELL AS A ALTERNATE CARTRIDGE OF PHBR SLIDES TO RESOLVE THE ISSUE. FOLLOWING THIS ACTION, ACCEPTABLE VITROS PHBR QUALITY CONTROL PERFORMANCE WAS OBSERVED. THE ROOT CAUSE OF THIS EVENT IS AN ATYPICAL CALIBRATION OF AN UNKNOWN CAUSE. THE CALIBRATOR FLUIDS, PHBR SLIDE CARTRIDGE IN USE AT THE TIME OF THE EVENT, OR AN ANALYZER RELATED ISSUE, COULD NOT BE RULED OUT AS A POTENTIAL CONTRIBUTING FACTORS OF THE EVENT.
A CUSTOMER OBSERVED A HIGHER THAN EXPECTED VITROS PHBR QUALITY CONTROL RESULTS ON A VITROS 5600 INTEGRATED SYSTEM. A RESULT OF 78.7 G/ML WAS OBTAINED FROM THE VITROS TDM PV CONTROL FLUID VERSUS THE EXPECTED VALUE OF 55.94 G/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF PATIENT SAMPLES WERE AFFECTED. PATIENT SAMPLES WERE NOT TESTED FOR VITROS PHBR WHILE QUALITY CONTROL RESULTS WERE OUTSIDE OF EXPECTED RANGES. THERE WAS NO REPORT OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITROS 5600 INTEGRATED SYSTEM | CHEMISTRY ANALYZER | DLZ | ORTHO-CLINICAL DIAGNOSTICS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |