FREESTYLE LIBRE 3
Report
- Report Number
- 2954323-2026-72391
- Event Type
- Malfunction
- Date Received
- May 14, 2026
- Date of Event
- February 7, 2024
- Report Date
- May 14, 2026
- Manufacturer
- ABBOTT DIABETES CARE LTD
- Product Code
- QBJ
- UDI-DI
- 05021791002658
- PMA / PMN Number
- K223435
- Removal / Correction Number
- 2954323-11/22/25-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
ADC INVESTIGATED THIS ISSUE AND DETERMINED THAT THE SENSOR ASSOCIATED WITH THIS COMPLAINT MAY NOT MEET PRODUCT DESIGN SPECIFICATIONS AND MAY PRODUCE LOW GLUCOSE READINGS WHICH ARE NOT CLINICALLY ACCEPTABLE. THIS PRODUCT HAS BEEN DETERMINED TO BE WITHIN THE SCOPE OF THE INVESTIGATION AND ACTIONS CONDUCTED UNDER ADC FIELD ACTION FA1002-2025. NO FURTHER INVESTIGATION ACTIONS ARE REQUIRED AS THIS ISSUE IS CONFIRMED TO BE IN THE SCOPE OF ADC FA1002-2025. HTTPS://EMDR.FDA.GOV/EMDR/FORMREDACTION/D11.JSF# ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
A CUSTOMER REPORTED RECEIVING LOW GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE (ADC) DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY ADC FIELD ACTION FA1002-2025. IMPACTED PRODUCT WAS DISTRIBUTED TO ANDORRA, AUSTRIA, BELGIUM, CANADA, DENMARK, FINLAND, FRANCE, GERMANY, ITALY, LUXEMBOURG, NETHERLANDS, NEW ZEALAND, NORWAY, SAN MARINO, SPAIN, SWEDEN, SWITZERLAND, UK, US, AND QATAR. ADC FIELD ACTION FA1002-2025 WAS ISSUED ONLY TO IMPACTED COUNTRIES. THIS REPORT IS BEING SUBMITTED AS PART OF RETROSPECTIVE REPORTING RELATED TO FA1002-2025. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 553995 | FREESTYLE LIBRE 3 | CONTINUOUS GLUCOSE MONITORING SYSTEM | QBJ | ABBOTT DIABETES CARE LTD | 72081-01 | T60001558 | 05021791002658 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |