PRESTIGE CERVICAL DISC SYSTEM
Report
- Report Number
- 1030489-2012-00407
- Event Type
- Injury
- Date Received
- April 4, 2012
- Report Date
- March 7, 2012
- Manufacturer
- WARSAW ORTHOPEDICS
- Product Code
- MJO
- PMA / PMN Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS DEVICE WAS USED DURING THE (B)(4) STUDY PRIOR TO DEVICE APPROVAL. THE APPROVED DEVICE PRODUCT CODE IS MJO AND THE PMA NUMBER IS P060018. (B)(4). THE DEVICE OR APPLICABLE IMAGING STUDIES HAVE NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. WE ARE UNABLE TO DETERMINE CAUSE OF THE EVENT.
IT WAS REPORTED THAT THE PATIENT UNDERWENT CERVICAL DISCECTOMY AND IMPLANT OF AN ARTIFICIAL CERVICAL DISC AT C5/6. APPROXIMATELY 42 MONTHS POSTOPERATIVELY, STUDY X-RAYS DEMONSTRATED SOME SETTLING OF THE DISC SPACE WITH SOME SMALL AMOUNT OF LUCENCY JUST ABOVE THE DISC SPACE ADJACENT TO THE IMPLANT. NO TREATMENT WAS REQUIRED. APPROXIMATELY 58 MONTHS POSTOPERATIVELY, THE PATIENT HAD NO SYMPTOMS, NORMAL RANGE OF MOTION, AND THE IMPLANT WAS IN GOOD POSITION. APPROXIMATELY 83 MONTHS POSTOPERATIVELY, DURING A FOLLOW-UP APPOINTMENT, IT WAS NOTED THAT THE SCREWS INFERIOR TO THE ARTIFICIAL DISC WERE BROKEN IN THE C6 VERTEBRAL BODY. NO TREATMENT WAS GIVEN. THE STUDY SITE NOTED THAT THE LUCENCY ABOVE THE DISC SPACE ADJACENT TO THE IMPLANT REMAINED UNCHANGED. NO TREATMENT WAS GIVEN. APPROXIMATELY 84 MONTHS POSTOPERATIVELY, A MYELOGRAM SHOWED FRACTURED C6 FIXATION SCREWS AND OSSIFICATION PRESENT AT THE POSTERIOR MARGIN OF THE C5-C6 INTERSPACE. THE PATIENT UNDERWENT REVISION SURGERY WHICH INCLUDED EXPLANT OF THE ARTIFICIAL DISC, ANTERIOR DISCECTOMY AND FUSION AT C5-C6 WITH ALLOGRAFT AND TITANIUM ANTERIOR PLATE, POSTERIOR LATERAL MASS SCREW INSTRUMENTATION, AND C5-C6 POSTEROLATERAL ARTHRODESIS WITH DBM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRESTIGE CERVICAL DISC SYSTEM | PROSTHESIS, INTERVERTEBRAL DISC | MJO | WARSAW ORTHOPEDICS | NA | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |