- Manufacturer
- ALCON - HOUSTON
- Product Code
- KYG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- NURSE
Narratives
KNIFE EVALUATION: NO SAMPLES WERE RETURNED FOR EVALUATION; THEREFORE, THE CONDITION OF THE PRODUCT COULD NOT BE VERIFIED. THE DEVICE HISTORY RECORD (DHR) FOR THE LOT WAS REVIEWED. FUNCTIONAL PENETRATION AND SHARPNESS TEST VALUES FOR THE LOT MET SPECIFICATION. NO ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO A DULL KNIFE WERE FOUND DURING THE DHR REVIEW AND THE PRODUCT WAS RELEASED ACCORDING TO COMPANY ACCEPTANCE CRITERIA. THE ROOT CAUSE COULD NOT BE DETERMINED. ALL KNIVES ARE 100 PERCENT INSPECTED BY TRAINED OPERATORS USING A MINIMUM OF 10X MAGNIFICATION DURING MANUFACTURING. PENETRATION AND SHARPNESS TESTING ARE PERFORMED AND MONITORED DURING THE FINISHING PROCESS TO ENSURE THE SHARPNESS OF THE PRODUCT. CUSTOM PAK EVALUATION: A SAMPLE HAS BEEN RECEIVED. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. (B)(4).
A NURSE REPORTED THAT DURING A SURGICAL PROCEDURE, THE KNIFE WAS TOO DULL TO ENTER THE PATIENT'S EYE. THE KNIFE WAS EXCHANGED FOR A NEW ONE AND THE PROCEDURE WAS COMPLETED. THERE WAS NO HARM TO THE PATIENT. IT IS ALSO NOTED THAT DURING THE SURGERY, THE DRAIN BAG WAS LEAKING. THERE WAS NO IMPACT ON THE SURGERY, AND NO HARM TO THE PATIENT, AS A RESULT OF THIS LEAK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CUSTOM PAK | CONVENIENCE KIT | KYG | ALCON - HOUSTON | CUSTOM PAK | 1252909H |