FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (HS-TNI ) CARTRIDGE

MDR report key: 25159195 · Received May 13, 2026

Report

Report Number
2245578-2026-00103
Event Type
Malfunction
Date Received
May 13, 2026
Date of Event
April 21, 2026
Report Date
May 13, 2026
Manufacturer
ABBOTT POINT OF CARE INC.
Product Code
MMI
UDI-DI
00054749003051
PMA / PMN Number
K240984
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 0

ON 21-APR-2026, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT FALSE POSITIVE RESULT ON A 57 YEAR OLD PATIENT. THERE WAS NO PATIENT INFORMATION, NO DETAILS SURROUNDING THE EVENT.. THERE WERE MULTIPLE REQUESTS FOR INFORMATION BUT TO NO AVAIL. METHOD: RESULT: SAMPLE: I-STAT, >1000 NG/L, WB A; I-STAT, >1000 NG/L, PLASMA A; I-STAT, >1000 NG/L, PLASMA A; LAB, 6 NG/L, PLASMA A; I-STAT, >1000 NG/L, PLASMA B; LAB, 5 NG/L, PLASMA B; LAB, 6 NG/L, PLASMA C; LAB, 6 NG/L, PLASMA D; I-STAT, 31.4 NG/L, WB E; I-STAT, 21.3 NG/L, PLASMA E; LAB, <4 NG/L, PLASMA E; I-STAT, 5.2 NG/L, PLASMA F; LAB, <4 NG/L, PLASMA F; NO TEST/COLECTION TIMES OR DATES PROVIDED. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE INVESTIGATION IS UNDERWAY. 99TH PERCENTILE 90% CI SEX N (NG/L, PG/ML) (NG/L, PG/ML) FEMALE 490 13 (10, 17) MALE 404 28 (19, 58) OVERALL 895 21 (14, 30).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
75979 I-STAT TROPONIN (HS-TNI ) CARTRIDGE HS-TNI CARTRIDGE MMI ABBOTT POINT OF CARE INC. NA NI 00054749003051

Patients

Seq Age Sex Outcome Treatment
1