FDA Adverse Event Malfunction Summary report: N

ABL90 FLEX PLUS ANALYZER

MDR report key: 25158520 · Received May 13, 2026

Report

Report Number
3002807968-2026-00058
Event Type
Malfunction
Date Received
May 13, 2026
Date of Event
March 29, 2026
Report Date
May 13, 2026
Manufacturer
RADIOMETER MEDICAL APS
Product Code
CHL
UDI-DI
05700693930923
PMA / PMN Number
K160153
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

RADIOMETER COMPLAINT REFERENCE: (B)(4). ACCORDING TO COMPLAINT, CUSTOMER REPORTED THAT THERE HAVE BEEN 5 DIFFERENT PATIENT SAMPLES THAT HAVE HAD VERY LOW TOTAL HEMOGLOBIN (THB) RESULTS RANGING FROM 0.1 TO 12 G/L BETWEEN (B)(6) 2026 FROM BLOOD GAS ANALYZER ABL90 FLEX PLUS ANALYZER (SERIAL NUMBER (B)(6)). THERE WERE NO ERROR MESSAGES ON THE SAMPLES AND SAMPLES FROM THESE PATIENTS PERFORMED IN THE LAB HAVE GIVEN NORMAL RESULTS. THE SAMPLE MEASUREMENTS ARE AS FOLLOWS: SAMPLE ID: MEASUREMENT TIME AND DATE: PARAMETER: RESULT: (B)(6) 22:09 (B)(6) 2026 THB, 1.4 G/L DISCREPANT, LAB RESULT 05:30 (B)(6) 2026 THB, 96 G/L COMPARISON, LAB RESULT 22:16 (B)(6) 2026 THB, 96 G/L COMPARISON, (B)(6) 22:11 (B)(6) 2026 THB, 70 G/L COMPARISON; (B)(6) 14:44 (B)(6) 2026 THB, 0.3 G/L DISCREPANT, LAB RESULT 07:40 (B)(6) 2026 THB, 86 G/L COMPARISON, LAB RESULT 07:38 (B)(6) 2026 THB, 90 G/L COMPARISON, (B)(6) 13:46 (B)(6) 2026 THB, 85 G/L COMPARISON, (B)(6) 09:34 (B)(6) 2026 THB, 84 G/L COMPARISON; (B)(6) 11:08 (B)(6) 2026 THB, 0.1 G/L DISCREPANT, LAB RESULT 09:09 (B)(6) 2026 THB, 108 G/L COMPARISON, LAB RESULT 14:31 (B)(6) 2026 THB, 102 G/L COMPARISON, (B)(6) 17:55 (B)(6) 2026 THB, 107 G/L COMPARISON, (B)(6) 15:14 (B)(6) 2026 THB, 103 G/L COMPARISON; (B)(6) 16:28 (B)(6) 2026 THB, 1.8 G/L DISCREPANT, LAB RESULT 08:58 (B)(6) 2026 THB, 104 G/L COMPARISON, LAB RESULT 13:38 (B)(6) 2026 THB, 114 G/L COMPARISON; (B)(6) 23:08 (B)(6) 2026 THB, 12 G/L DISCREPANT, LAB RESULT 08:41 (B)(6) 2026 THB, 85 G/L COMPARISON, LAB RESULT 22:51 (B)(6) 2026 THB, 92 G/L COMPARISON, (B)(6) 15:04 (B)(6) 2026 THB, 89 G/L COMPARISON, (B)(6) 03:41 (B)(6) 2026 THB, 90 G/L COMPARISON.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
302679 ABL90 FLEX PLUS ANALYZER ABL90 FLEX PLUS ANALYZER CHL RADIOMETER MEDICAL APS 393-092 05700693930923

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown