FDA Adverse Event Malfunction Summary report: N

BRAIN MONITOR WHT-REFURB

MDR report key: 25157951 · Received May 13, 2026

Report

Report Number
9612330-2026-00001
Event Type
Malfunction
Date Received
May 13, 2026
Date of Event
April 14, 2026
Report Date
May 13, 2026
Manufacturer
NATUS MEDICAL INCORPORATED DBA EXCEL-TECH LTD.
Product Code
OLV
UDI-DI
00382830002523
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

FOLLOW UP REPORT 001 REFERENCE NATUS COMPLAINT# (B)(4). PER (B)(4) REV 87 XLTEK EEG PSG RAS. HAZARD 2.1 - FAILURE OF AMPLIFIER POWER SUPPLY/ TRANSFORMER. EFFECT (HARM): FIRE - SMOKE INHALATION, SECOND DEGREE BURNS, THIRD DEGREE BURNS. RESIDUAL RISK: MEDIUM. THE HAZARDS IDENTIFIED HAVE BEEN EVALUATED AND FOUND TO BE IN COMPLIANCE WITH A KNOWN STANDARD. AS NOTED IN QMS-000018, HAZARDS THAT HAVE BEEN EVALUATED AND FOUND TO BE IN COMPLIANCE WITH KNOWN STANDARDS CAN BE PRESUMED TO BE CONSISTENT WITH AN ACCEPTABLE LEVEL OF RISK, YIELDING AN ACCEPTABLE RISK / BENEFIT TO THE PATIENT OR USER. RISK OUTWEIGHED BY BENEFIT OF USE OF DEVICE. INVESTIGATION RESULTS: THE POWER CORD PLUG WAS EVALUATED BY NATUS ENGINEERING BY VISUAL INSPECTION; THE FAILURE APPEARS TO BE RELATED TO WEAR AND TEAR. THE POWER CORD PLUG IS AN OLDER POWERVAR DESIGN, WHICH WAS ENHANCED IN 2024 FROM AN ASSEMBLE PLUG TO A MOLDED PLUG, MAKING THE CORD MORE ROBUST. ROOT CAUSE/PROBABLE ROOT CAUSE: WEAR AND TEAR, THE PLUG USED IN THIS INSTANCE WAS THE PRIOR ASSEMBLED PLUG DESIGN AND IS BEYOND THE 5-YEAR WARRANTY PERIOD PROVIDED BY THE MANUFACTURER.

Additional Manufacturer Narrative · 0

INITIAL REPORT REFERENCE NATUS COMPLAINT# (B)(4). PER (B)(4) REV 87 XLTEK EEG PSG RAS. HAZARD 2.1 - FAILURE OF AMPLIFIER POWER SUPPLY/ TRANSFORMER. EFFECT (HARM): FIRE - SMOKE INHALATION, SECOND DEGREE BURNS, THIRD DEGREE BURNS. RESIDUAL RISK: MEDIUM THE HAZARDS IDENTIFIED HAVE BEEN EVALUATED AND FOUND TO BE IN COMPLIANCE WITH A KNOWN STANDARD. AS NOTED IN QMS-000018, HAZARDS THAT HAVE BEEN EVALUATED AND FOUND TO BE IN COMPLIANCE WITH KNOWN STANDARDS CAN BE PRESUMED TO BE CONSISTENT WITH AN ACCEPTABLE LEVEL OF RISK, YIELDING AN ACCEPTABLE RISK / BENEFIT TO THE PATIENT OR USER. RISK OUTWEIGHED BY BENEFIT OF USE OF DEVICE. THE CUSTOMER REPORTED THE POWER CORD CAUGHT FIRE. THERE WAS NO INJURY, NO ENVIRONMENTAL HAZARD. THE CUSTOMER PROVIDED PICTURES OF THE AFFECTED PART. THE UNIT IS END OF LIFE. THE CUSTOMER HAS BEEN REQUESTED TO RETURN THE UNIT FOR EVALUATION. FURTHER INVESTIGATION TO BE CARRIED OUT.

Description of Event or Problem · 0

POWER CORD/PLUG FROM THE UPS CAUGHT FIRE. IT HAS SCORCH MARKS AND MELTING AROUND THE CORD. NO INJURIES.

Description of Event or Problem · 0

POWER CORD/PLUG FROM THE UPS CAUGHT FIRE. IT HAS SCORCH MARKS AND MELTING AROUND THE CORD. NO INJURIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
137744 BRAIN MONITOR WHT-REFURB BRAIN MONITOR WHT-REFURB OLV NATUS MEDICAL INCORPORATED DBA EXCEL-TECH LTD. 10388-R 00382830002523

Patients

Seq Age Sex Outcome Treatment
1