FDA Adverse Event Malfunction Summary report: N

STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL

MDR report key: 2515381 · Received April 3, 2012

Report

Report Number
3005099803-2012-01093
Event Type
Malfunction
Date Received
April 3, 2012
Date of Event
December 14, 2011
Report Date
December 14, 2011
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
FGO
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(6). NOTE: THIS EVENT, AS REPORTED, DID NOT REFLECT AN MDR REPORTABLE SCENARIO; HOWEVER, EVALUATION OF THE RETURNED DEVICE REVEALED AN MDR-REPORTABLE MALFUNCTION. THIS MANUFACTURER REPORT IS BEING SUBMITTED BASED ON THE EVALUATION RESULTS. ANALYSIS OF THE RETURNED STONE CONE RETRIEVAL COIL REVEALED A KINK IN THE BLUE SHEATH. FURTHER VISUAL INSPECTION NOTED THAT THE BLUE/GREEN HEAT SHRINK WAS MISSING FROM DISTAL STOP TO THE DISTAL BALL. FUNCTIONAL EVALUATION WAS PERFORMED AND THE BASKET WOULD NOT FULLY CLOSE. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED AND CONFIRMED THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH THE DEVICE MASTER RECORD. THE DEFECTS IDENTIFIED ON THE DEVICE WERE MOST LIKELY DUE TO SOME OPERATIONAL OR ANATOMICAL ASPECT ENCOUNTERED DURING THE PROCEDURE; THEREFORE, THE MOST PROBABLE ROOT CAUSE FOR THIS COMPLAINT IS OPERATIONAL/PHYSIOLOGICAL CONTEXT.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A STONE CONE RETRIEVAL COIL WAS USED IN A PROCEDURE ON (B)(6) 2011. THE PATIENT WAS MALE (IDENTIFIER, AGE, DATE OF BIRTH AND WEIGHT UNKNOWN). ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE WHILE INSIDE THE PATIENT, THE DEVICE WAS UNABLE TO OPEN AND SEVER RESISTANCE WAS MET. AFTER REMOVING THE DEVICE FROM THE PATIENT, THE DEVICE WAS TESTED AND IT WAS ABLE TO OPEN AND CLOSE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION POST PROCEDURE WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL DISLODGER, STONE, FLEXIBLE FGO BOSTON SCIENTIFIC - MARLBOROUGH M0063903200 0000895450

Patients

Seq Age Sex Outcome Treatment
1