STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL
Report
- Report Number
- 3005099803-2012-01093
- Event Type
- Malfunction
- Date Received
- April 3, 2012
- Date of Event
- December 14, 2011
- Report Date
- December 14, 2011
- Manufacturer
- BOSTON SCIENTIFIC - MARLBOROUGH
- Product Code
- FGO
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(6). NOTE: THIS EVENT, AS REPORTED, DID NOT REFLECT AN MDR REPORTABLE SCENARIO; HOWEVER, EVALUATION OF THE RETURNED DEVICE REVEALED AN MDR-REPORTABLE MALFUNCTION. THIS MANUFACTURER REPORT IS BEING SUBMITTED BASED ON THE EVALUATION RESULTS. ANALYSIS OF THE RETURNED STONE CONE RETRIEVAL COIL REVEALED A KINK IN THE BLUE SHEATH. FURTHER VISUAL INSPECTION NOTED THAT THE BLUE/GREEN HEAT SHRINK WAS MISSING FROM DISTAL STOP TO THE DISTAL BALL. FUNCTIONAL EVALUATION WAS PERFORMED AND THE BASKET WOULD NOT FULLY CLOSE. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED AND CONFIRMED THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH THE DEVICE MASTER RECORD. THE DEFECTS IDENTIFIED ON THE DEVICE WERE MOST LIKELY DUE TO SOME OPERATIONAL OR ANATOMICAL ASPECT ENCOUNTERED DURING THE PROCEDURE; THEREFORE, THE MOST PROBABLE ROOT CAUSE FOR THIS COMPLAINT IS OPERATIONAL/PHYSIOLOGICAL CONTEXT.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A STONE CONE RETRIEVAL COIL WAS USED IN A PROCEDURE ON (B)(6) 2011. THE PATIENT WAS MALE (IDENTIFIER, AGE, DATE OF BIRTH AND WEIGHT UNKNOWN). ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE WHILE INSIDE THE PATIENT, THE DEVICE WAS UNABLE TO OPEN AND SEVER RESISTANCE WAS MET. AFTER REMOVING THE DEVICE FROM THE PATIENT, THE DEVICE WAS TESTED AND IT WAS ABLE TO OPEN AND CLOSE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION POST PROCEDURE WAS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL | DISLODGER, STONE, FLEXIBLE | FGO | BOSTON SCIENTIFIC - MARLBOROUGH | M0063903200 | 0000895450 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |