CADD CLEO INFUSION SET
Report
- Report Number
- 3012307300-2026-04840
- Event Type
- Malfunction
- Date Received
- May 11, 2026
- Date of Event
- April 16, 2026
- Report Date
- May 12, 2026
- Manufacturer
- SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 10610586028335
- PMA / PMN Number
- K042172
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- 003
Narratives
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
PATIENT WITH DIABETES STATED THAT GLUCOSE READINGS WERE HIGH AROUND 9:00 PM. PWD CHANGED THE CANNULA, AFTER WHICH A LINE BLOCKED ALARM WAS RECEIVED. PWD THEN CHANGED BOTH THE CANNULA AND TUBING. AT APPROXIMATELY 2:00 AM, PWD REPORTED THAT GLUCOSE LEVELS WERE STILL HIGH, PROMPTING A CHANGE OF THE CASSETTE AND INFUSION SET. PWD ADMINISTERED A BOLUS TO TREAT THE HIGH GLUCOSE AND WAITED TO SEE IF LEVELS WOULD DECREASE; HOWEVER, GLUCOSE DID NOT DROP. PWD THEN USED INHALABLE INSULIN, WHICH BEGAN TO BRING GLUCOSE LEVELS DOWN. PWD REPORTED GLUCOSE LEVELS IN THE 200 MG/DL RANGE. PER HALO, GLUCOSE WAS READING 185 MG/DL. PWD STATED THAT A LINE BLOCKED ALARM WAS ALSO EXPERIENCED THE PREVIOUS WEEK. PWD FOLLOWED ON-SCREEN TROUBLESHOOTING STEPS, CHECKED THE TUBING (NO KINKS NOTED), AND CHANGED THE INFUSION SITE. DESPITE REPEATING TROUBLESHOOTING STEPS THREE TO FOUR TIMES, THE ALARM PERSISTED. PWD THEN CHANGED THE CASSETTE, WHICH RESOLVED THE ALARM. PWD REQUESTED REPLACEMENTS. ADDRESS WAS CONFIRMED. PSS OFFERED TO SEND AN SMS TO COLLECT LOT NUMBERS FOR THE CASSETTES AND INFUSION SETS, AND PWD AGREED. THE EVENT OCCURRED WHILE IN USE WITH PATIENT. THE OPERATOR OF THE DEVICE WAS THE LAY USER OR PATIENT. THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 452956 | CADD CLEO INFUSION SET | SET, ADMINISTRATION, INTRAVASCULAR | FPA | SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V. | 10610586028335 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |