FDA Adverse Event Malfunction Summary report: N

ANTI-C (MONOCLONAL)

MDR report key: 2514139 · Received April 3, 2012

Report

Report Number
1034569-2012-00064
Event Type
Malfunction
Date Received
April 3, 2012
Date of Event
March 9, 2012
Report Date
March 30, 2012
Manufacturer
IMMUCOR, INC.
Product Code
KSZ
PMA / PMN Number
STN 103498
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

THE IMAGE RESULT FILES WERE REVIEWED BY AN IMMUCOR TECHNICAL SUPPORT SPECIALIST. SAMPLE-(B)(4) RESULTED AS (B)(4) ON THE ECHO AS (WELL APPEARS POSITIVE WITH VERY WEAK AGGLUTINATION) HOWEVER CUSTOMER STATES THIS IS A KNOWN (B)(4) SAMPLE. THE SAMPLE WAS REPEATED WITH A NEW LOT OF ANTI-C (LOT# 936061). POSITIVE (B)(4) REACTIONS WERE OBTAINED. RESULTS VISUALLY APPEARED TO BE LOOSELY SWIRLED AGGLUTINATION. POSITIVE (B)(4) REACTIONS WERE OBTAINED. RESULTS VISUALLY APPEARED TO BE LOOSELY SWIRLED AGGLUTINATION. CUSTOMER WAS ADVISED TO PERFORM A FULL RH PHENOTYPE ON THE SAMPLE AND TO ALSO RETEST FOR THE (B)(4) ANTIGEN IN TUBE ALONGSIDE A KNOWN C ANTIGEN (B)(4) CONTROL TO STANDARDIZE THE SHAKING OF TUBES. SAMPLE (B)(4) WAS LITTLE (B)(4), BIG (B)(4), AND LITTLE (B)(4). TUBE TESTING FOR BIG (B)(4). CUSTOMER WAS ASKED TO REPEAT TESTING ON THE ECHO. (B)(4) RESULTS FOR THE (B)(4) ANTIGEN WERE OBTAINED. CUSTOMER WAS MADE AWARE THAT HETEROZYGOUS SAMPLES WILL HAVE WEAKER REACTIVITY. IN-HOUSE TESTING WAS PERFORMED ON RETENTION PRODUCTS. THE AG_C RH2 ASSAY WAS PERFORMED ON TWO (B)(4) ANTIGEN IN-HOUSE DONORS USING RETENTION CMT PLATES, LOT NU514 AND ANTI-C, LOT 936060 ON THE ECHO. CONTROLS PERFORMED AS EXPECTED AND IN-HOUSE DONORS RESULTED (B)(4) AS EXPECTED. PERFORMED HEMAGGLUTINATION TUBE TESTING ON TWO (B)(4) ANTIGEN IN-HOUSE DONORS USING RETENTION ANTI-C, LOT 936060. CONTROLS PERFORMED AS EXPECTED AND IN-HOUSE DONORS EXHIBITED (B)(4) REACTIVITY. RETENTION PRODUCT PERFORMED AS EXPECTED. ALL ECHO CUSTOMERS WERE NOTIFIED THAT NEGATIVE ANTIBODY SCREENING AND IDENTIFICATION RESULTS ON THE ECHO ARE TO BE VISUALLY INSPECTED PRIOR TO RELEASE OF RESULTS. THIS ISSUE WAS COMMUNICATED TO CUSTOMERS IN TECHNICAL COMMUNICATION (B)(4) ON (B)(4), 2009.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT UNEXPECTED NEGATIVE REACTIONS WERE OBTAINED WHILE PERFORMING C ANTIGEN TESTING ON THE ECHO INSTRUMENT USING BLOOD GROUPING REAGENT ANTI-C (MONOCLONAL) GAMMA-CLONE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANTI-C (MONOCLONAL) BLOOD GROUPING REAGENT KSZ IMMUCOR, INC. 936060

Patients

Seq Age Sex Outcome Treatment
1