ANTI-C (MONOCLONAL)
Report
- Report Number
- 1034569-2012-00064
- Event Type
- Malfunction
- Date Received
- April 3, 2012
- Date of Event
- March 9, 2012
- Report Date
- March 30, 2012
- Manufacturer
- IMMUCOR, INC.
- Product Code
- KSZ
- PMA / PMN Number
- STN 103498
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
THE IMAGE RESULT FILES WERE REVIEWED BY AN IMMUCOR TECHNICAL SUPPORT SPECIALIST. SAMPLE-(B)(4) RESULTED AS (B)(4) ON THE ECHO AS (WELL APPEARS POSITIVE WITH VERY WEAK AGGLUTINATION) HOWEVER CUSTOMER STATES THIS IS A KNOWN (B)(4) SAMPLE. THE SAMPLE WAS REPEATED WITH A NEW LOT OF ANTI-C (LOT# 936061). POSITIVE (B)(4) REACTIONS WERE OBTAINED. RESULTS VISUALLY APPEARED TO BE LOOSELY SWIRLED AGGLUTINATION. POSITIVE (B)(4) REACTIONS WERE OBTAINED. RESULTS VISUALLY APPEARED TO BE LOOSELY SWIRLED AGGLUTINATION. CUSTOMER WAS ADVISED TO PERFORM A FULL RH PHENOTYPE ON THE SAMPLE AND TO ALSO RETEST FOR THE (B)(4) ANTIGEN IN TUBE ALONGSIDE A KNOWN C ANTIGEN (B)(4) CONTROL TO STANDARDIZE THE SHAKING OF TUBES. SAMPLE (B)(4) WAS LITTLE (B)(4), BIG (B)(4), AND LITTLE (B)(4). TUBE TESTING FOR BIG (B)(4). CUSTOMER WAS ASKED TO REPEAT TESTING ON THE ECHO. (B)(4) RESULTS FOR THE (B)(4) ANTIGEN WERE OBTAINED. CUSTOMER WAS MADE AWARE THAT HETEROZYGOUS SAMPLES WILL HAVE WEAKER REACTIVITY. IN-HOUSE TESTING WAS PERFORMED ON RETENTION PRODUCTS. THE AG_C RH2 ASSAY WAS PERFORMED ON TWO (B)(4) ANTIGEN IN-HOUSE DONORS USING RETENTION CMT PLATES, LOT NU514 AND ANTI-C, LOT 936060 ON THE ECHO. CONTROLS PERFORMED AS EXPECTED AND IN-HOUSE DONORS RESULTED (B)(4) AS EXPECTED. PERFORMED HEMAGGLUTINATION TUBE TESTING ON TWO (B)(4) ANTIGEN IN-HOUSE DONORS USING RETENTION ANTI-C, LOT 936060. CONTROLS PERFORMED AS EXPECTED AND IN-HOUSE DONORS EXHIBITED (B)(4) REACTIVITY. RETENTION PRODUCT PERFORMED AS EXPECTED. ALL ECHO CUSTOMERS WERE NOTIFIED THAT NEGATIVE ANTIBODY SCREENING AND IDENTIFICATION RESULTS ON THE ECHO ARE TO BE VISUALLY INSPECTED PRIOR TO RELEASE OF RESULTS. THIS ISSUE WAS COMMUNICATED TO CUSTOMERS IN TECHNICAL COMMUNICATION (B)(4) ON (B)(4), 2009.
CUSTOMER REPORTED THAT UNEXPECTED NEGATIVE REACTIONS WERE OBTAINED WHILE PERFORMING C ANTIGEN TESTING ON THE ECHO INSTRUMENT USING BLOOD GROUPING REAGENT ANTI-C (MONOCLONAL) GAMMA-CLONE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANTI-C (MONOCLONAL) | BLOOD GROUPING REAGENT | KSZ | IMMUCOR, INC. | 936060 |