Narratives
H6 INVESTIGATION FINDINGS C19 REFERS TO THE PRODUCT HISTORY REVIEW: A REVIEW OF THE MANUFACTURING RECORDS FOR THE DEVICE VERIFIED THAT THE LOT MET ALL PRERELEASE SPECIFICATIONS. CBAS® HEPARIN SURFACE INCORPORATES CARMEDA HEPARIN MANUFACTURED FROM HEPARIN SODIUM API, WHICH IS COVALENTLY BOUND TO THE DEVICE SURFACE AND IS ESSENTIALLY NON-ELUTING. W. L. GORE & ASSOCIATES, INC. (GORE) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY GORE, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. BLANK FIELDS PRESENT ON THIS REPORT INCLUDE REQUIRED FIELDS AND FIELDS DETERMINED TO BE NOT APPLICABLE. BLANK REQUIRED FIELDS INDICATE THAT THE INFORMATION WAS NOT PROVIDED, WAS DEEMED UNAVAILABLE OR WAS NOT APPLICABLE. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, GORE, OR ITS ASSOCIATES THAT THE DEVICE, GORE OR ITS ASSOCIATES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE A LEGAL ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY DEFECTS OR HAS MALFUNCTIONED, AS DEFINED FROM A LEGAL STANDPOINT. THESE WORDS ARE INCLUDED IN THE REPORT AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
THE FOLLOWING WAS REPORTED TO GORE FROM THE VBX 21-04 MEDRIO STUDY DATABASE: ON (B)(6) 2021, AN 80-YEAR-OLD PATIENT UNDERWENT ENDOVASCULAR TREATMENT FOR THORACOABDOMINAL ANEURYSM III. THE PATIENT WAS TREATED WITH A JOTEC E-NSIDE BRANCHED GRAFT AND FOUR GORE®VIABAHN®VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX STENTS). ONE VBX STENT WAS PLACED IN THE CELIAC TRUNK, ONE IN THE SUPERIOR MESENTERIC ARTERY (SMA) AND THE OTHER TWO VBX STENTS WERE PLACED IN THE BILATERAL RENAL ARTERIES. THE FOUR VBX STENTS WERE SUCCESSFULLY IMPLANTED AND WERE NOTED AS PATENT AT THE END OF THE PROCEDURE. ACCORDING TO THE REPORT, ON (B)(6) 2025, KINKING WITHOUT STENOSIS OF THE VBX STENT IN SMA DUE TO THE MIGRATION OF THE BRANCHED JOTEC E-NSIDE GRAFT WAS OBSERVED. AN ENDOVASCULAR REPEAT INTERVENTION WAS PERFORMED ON (B)(6) 2025 AND RELINED THE VBX STENT WITH A DYNAMIC 8.0X38 BARE METAL STENT TO ADDRESS THE KINKING OF THE VBX STENT CAUSED FROM THE MIGRATION OF E-NSIDE GRAFT. THE VBX STENT IS REPORTED AS REMAINING IN PLACE AND NOT EXPLANTED. THE PATIENT RECOVERED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 173401 | GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS | ILIAC COVERED STENT, ARTERIAL | PRL | W. L. GORE & ASSOCIATES, INC. | 00733132620890 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Male | JOTEC E-NSIDE BRANCHED GRAFT: 38-26 65MU382621-4B8 |